Beta Bionics, Inc. — Class II recall: iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software
- Event date
- 2026-04-22 00:00:00Z
- Observed
- 2026-07-09 22:53:03Z
- Source ID
- Z-1809-2026
- Payload hash
- 365ad600926c05d90e4e6973…
Summary
it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar… Status: Ongoing.