fda_recall · device_recall · high

Centinel Spine, Inc. — Class II recall: Prodisc C SK U.S. IMPLANT EXTRA LARGE 5MM. Model: PDSXL5 Total cervical disc…

Event date
2026-05-13 00:00:00Z
Observed
2026-07-09 22:52:35Z
Source ID
Z-2050-2026
Payload hash
1bde363bd691a87bdfececc9

Summary

Product labeling mix up. The prodisc C SK U.S. Implant Extra Large 6mm product was labeled as a 5mm and prodisc C SK U.S. Implant Extra Large 5mm product was labeled as a 6mm. Status: Ongoing.

Informational only. Not legal, financial, or medical advice.