fda_recall · device_recall · high

Dexcom, Inc. — Class II recall: Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom…

Event date
2026-05-20 00:00:00Z
Observed
2026-07-09 22:52:30Z
Source ID
Z-2163-2026
Payload hash
992c4a54063f8141fe2b1cb3

Summary

A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the ap… Status: Ongoing.

Entities