fda_recall · device_recall · high

Dexcom, Inc. — Class II recall: Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexc…

Event date
2026-05-20 00:00:00Z
Observed
2026-07-09 22:52:32Z
Source ID
Z-2165-2026
Payload hash
e484dc76f6a125dce512ad11

Summary

A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the ap… Status: Ongoing.

Entities