fda_recall · device_recall · high

GE Medical Systems Information Technologies Inc — Class II recall: GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL, REF #2063702-101, Product…

Event date
2026-06-03 00:00:00Z
Observed
2026-07-09 22:52:25Z
Source ID
Z-2242-2026
Payload hash
17aa2beecc57f15ca44eb81c

Summary

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Cen… Status: Ongoing.

Informational only. Not legal, financial, or medical advice.