fda_recall · device_recall · high
GE Medical Systems Information Technologies Inc — Class II recall: GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL, REF #2063702-101, Product…
- Event date
- 2026-06-03 00:00:00Z
- Observed
- 2026-07-09 22:52:25Z
- Source ID
- Z-2242-2026
- Payload hash
- 17aa2beecc57f15ca44eb81c…
Summary
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Cen… Status: Ongoing.
Informational only. Not legal, financial, or medical advice.

