fda_recall · device_recall · high

GE Medical Systems, LLC — Class II recall: Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1, a combination…

Event date
2026-06-24 00:00:00Z
Observed
2026-07-09 22:52:09Z
Source ID
Z-2476-2026
Payload hash
e0e9fa014f37a2524d1962c6

Summary

GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into t… Status: Ongoing.

Informational only. Not legal, financial, or medical advice.