fda_recall · device_recall · high

Medline Industries, LP — Class II recall: Medline Convenience Kits: 1) ANCILLARY PACK-LF, Model Number:DYNJ0185290G; 2)…

Event date
2026-05-13 00:00:00Z
Observed
2026-07-09 22:52:40Z
Source ID
Z-2089-2026
Payload hash
f94e457cc812f67e711bf5a1

Summary

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL)… Status: Ongoing.

Informational only. Not legal, financial, or medical advice.