fda_recall · device_recall · high

The Binding Site Group, Ltd. — Class II recall: Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number:…

Event date
2026-06-17 00:00:00Z
Observed
2026-07-09 22:52:12Z
Source ID
Z-2362-2026
Payload hash
1fe8583ee967e1fcc523e45c

Summary

It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protei… Status: Ongoing.