Key facts
| Recalling firm | Erbe Medical, LLC |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF: 20325-207; ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Set for Fujifilm Scopes, REF: 20325-248; Hybrid CO2 Tubing/Cap Set for Fujinon Scopes, REF: 20325-208 |
| Classification | Class II |
| Recall number | Z-2062-2026 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure. |
What this classification means
FDA classifies a recall as Class II when using the recalled product, or being exposed to it, may cause temporary or medically reversible adverse health consequences, or when the probability of serious adverse health consequences is remote. In this case, the recalling firm is Erbe Medical, LLC and the affected product is described in FDA's record as: ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF (additional items listed in FDA's full record). FDA's stated reason for the recall is: Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to (additional items listed in FDA's full record). The recall is tracked under FDA recall number Z-2062-2026 and carries a status of Ongoing as of the most recent FDA enforcement report. This classification reflects FDA's assessment of the product defect described above, not a prediction about any specific patient outcome.
What consumers should do
Because this recall carries FDA's middle severity classification, anyone who has this product should stop using it and follow the recalling firm's instructions, even though serious harm is considered unlikely for this classification. Specifically, check whether the product in your possession matches the description on file with FDA -- ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF (additional items listed in FDA's full record) -- and the recall number Z-2062-2026. If it matches, stop using or distributing it and contact Erbe Medical, LLC or the point of purchase for return or replacement instructions. Retailers and distributors holding affected inventory should quarantine it and follow the firm's written recall notice.
Erbe Medical, LLC’s FDA history
Argus HQ has recorded 3 total FDA actions tied to Erbe Medical, LLC: 0 warning letters, 3 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- What is the FDA recall number for this event?
- FDA recall number Z-2062-2026.
- What is the current status of this recall?
- Status as recorded by FDA: Ongoing.
- Where can I verify this recall directly with FDA?
- The verbatim FDA record is the source of truth for every recall Argus HQ tracks — always confirm against fda.gov before acting.
- What was recalled and why?
- Erbe Medical, LLC recalled ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF (additional items listed in FDA's full record). FDA's record states the reason for recall as: Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Erbe Medical, LLCCompiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2026). Recall Record: Class II Recall: ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for… by Erbe Medical, LLC. Digital Empire LLC. Retrieved from https://argushq.ai/recall/fda-recall-erbe-medical-llc-z-2062-2026
"Recall Record: Class II Recall: ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for… by Erbe Medical, LLC." Argus HQ Research, Digital Empire LLC, 2026, argushq.ai/recall/fda-recall-erbe-medical-llc-z-2062-2026.
Argus HQ Research. "Recall Record: Class II Recall: ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for… by Erbe Medical, LLC." Digital Empire LLC. Accessed July 13, 2026. https://argushq.ai/recall/fda-recall-erbe-medical-llc-z-2062-2026.
@misc{argushq_argushq_ai_recall_fda_recall_erbe_medical_llc_z_2062_2026_2026,
title = {Recall Record: Class II Recall: ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for… by Erbe Medical, LLC},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/recall/fda-recall-erbe-medical-llc-z-2062-2026},
note = {Accessed: July 13, 2026}
}Source: FDA.gov — Recalls ↗