Key facts
| Recalling firm | Gentuity, LLC |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13; |
| Classification | Class II |
| Recall number | Z-1823-2026 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging. |
What consumers should do
Because this recall carries FDA's middle severity classification, anyone who has this product should stop using it and follow the recalling firm's instructions, even though serious harm is considered unlikely for this classification. Specifically, check whether the product in your possession matches the description on file with FDA -- Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13; -- and the recall number Z-1823-2026. If it matches, stop using or distributing it and contact Gentuity, LLC or the point of purchase for return or replacement instructions. Retailers and distributors holding affected inventory should quarantine it and follow the firm's written recall notice. FDA maintains a public enforcement report database so that consumers, physicians, and purchasers can look up the exact product identifiers involved before deciding what to do.
What this classification means
FDA classifies a recall as Class II when using the recalled product, or being exposed to it, may cause temporary or medically reversible adverse health consequences, or when the probability of serious adverse health consequences is remote. In this case, the recalling firm is Gentuity, LLC and the affected product is described in FDA's record as: Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;. FDA's stated reason for the recall is: The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain (additional items listed in FDA's full record). The recall is tracked under FDA recall number Z-1823-2026 and carries a status of Ongoing as of the most recent FDA enforcement report. This classification reflects FDA's assessment of the product defect described above, not a prediction about any specific patient outcome. FDA maintains a public enforcement report database so that consumers, physicians, and purchasers can look up the exact product identifiers involved before deciding what to do. Recall status can change while FDA and the recalling firm continue corrective actions, so checking the firm's own notice for the most current guidance is recommended.
Gentuity, LLC’s FDA history
Argus HQ has recorded 1 total FDA action tied to Gentuity, LLC: 0 warning letters, 1 recall, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- What is the FDA recall number for this event?
- FDA recall number Z-1823-2026.
- What is the current status of this recall?
- Status as recorded by FDA: Ongoing.
- Where can I verify this recall directly with FDA?
- The verbatim FDA record is the source of truth for every recall Argus HQ tracks — always confirm against fda.gov before acting.
- What was recalled and why?
- Gentuity, LLC recalled Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;. FDA's record states the reason for recall as: The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain (additional items listed in FDA's full record). The recall is logged under FDA recall number Z-1823-2026.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Gentuity, LLCCompiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2026). FDA Class II Recall: Gentuity HF-OCT Imaging System Model G10-01; Software version… by Gentuity, LLC — Recall Details. Digital Empire LLC. Retrieved from https://argushq.ai/recall/fda-recall-gentuity-llc-z-1823-2026
"FDA Class II Recall: Gentuity HF-OCT Imaging System Model G10-01; Software version… by Gentuity, LLC — Recall Details." Argus HQ Research, Digital Empire LLC, 2026, argushq.ai/recall/fda-recall-gentuity-llc-z-1823-2026.
Argus HQ Research. "FDA Class II Recall: Gentuity HF-OCT Imaging System Model G10-01; Software version… by Gentuity, LLC — Recall Details." Digital Empire LLC. Accessed July 13, 2026. https://argushq.ai/recall/fda-recall-gentuity-llc-z-1823-2026.
@misc{argushq_argushq_ai_recall_fda_recall_gentuity_llc_z_1823_2026_2026,
title = {FDA Class II Recall: Gentuity HF-OCT Imaging System Model G10-01; Software version… by Gentuity, LLC — Recall Details},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/recall/fda-recall-gentuity-llc-z-1823-2026},
note = {Accessed: July 13, 2026}
}Source: FDA.gov — Recalls ↗