Key facts
| Recalling firm | Physio-Control, Inc. |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122 |
| Classification | Class II |
| Recall number | Z-2065-2026 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy. |
What consumers should do
Because this recall carries FDA's middle severity classification, anyone who has this product should stop using it and follow the recalling firm's instructions, even though serious harm is considered unlikely for this classification. Specifically, check whether the product in your possession matches the description on file with FDA -- LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372 (additional items listed in FDA's full record) -- and the recall number Z-2065-2026. If it matches, stop using or distributing it and contact Physio-Control, Inc. or the point of purchase for return or replacement instructions. Retailers and distributors holding affected inventory should quarantine it and follow the firm's written recall notice.
What this classification means
FDA classifies a recall as Class II when using the recalled product, or being exposed to it, may cause temporary or medically reversible adverse health consequences, or when the probability of serious adverse health consequences is remote. In this case, the recalling firm is Physio-Control, Inc. and the affected product is described in FDA's record as: LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372 (additional items listed in FDA's full record). FDA's stated reason for the recall is: Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.. The recall is tracked under FDA recall number Z-2065-2026 and carries a status of Ongoing as of the most recent FDA enforcement report. This classification reflects FDA's assessment of the product defect described above, not a prediction about any specific patient outcome. FDA maintains a public enforcement report database so that consumers, physicians, and purchasers can look up the exact product identifiers involved before deciding what to do.
Physio-Control, Inc.’s FDA history
Argus HQ has recorded 1 total FDA action tied to Physio-Control, Inc.: 0 warning letters, 1 recall, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- What is the current status of this recall?
- Status as recorded by FDA: Ongoing.
- Where can I verify this recall directly with FDA?
- The verbatim FDA record is the source of truth for every recall Argus HQ tracks — always confirm against fda.gov before acting.
- What was recalled and why?
- Physio-Control, Inc. recalled LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372 (additional items listed in FDA's full record). FDA's record states the reason for recall as: Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.. The.
- How serious is this recall?
- FDA classifies a recall as Class II when using the recalled product, or being exposed to it, may cause temporary or medically reversible adverse health consequences, or when the probability of serious adverse health consequences is remote.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Physio-Control, Inc.Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2026). FDA Recall Database: Class II Recall: LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241,… by Physio-Control, Inc.. Digital Empire LLC. Retrieved from https://argushq.ai/recall/fda-recall-physio-control-inc-z-2065-2026
"FDA Recall Database: Class II Recall: LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241,… by Physio-Control, Inc.." Argus HQ Research, Digital Empire LLC, 2026, argushq.ai/recall/fda-recall-physio-control-inc-z-2065-2026.
Argus HQ Research. "FDA Recall Database: Class II Recall: LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241,… by Physio-Control, Inc.." Digital Empire LLC. Accessed July 13, 2026. https://argushq.ai/recall/fda-recall-physio-control-inc-z-2065-2026.
@misc{argushq_argushq_ai_recall_fda_recall_physio_control_inc_z_2065_2026_2026,
title = {FDA Recall Database: Class II Recall: LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241,… by Physio-Control, Inc.},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/recall/fda-recall-physio-control-inc-z-2065-2026},
note = {Accessed: July 13, 2026}
}Source: FDA.gov — Recalls ↗