fda_recall · device_recall · high
Accuray Incorporated — Class II recall: Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator…
- Event date
- 2026-07-08 00:00:00Z
- Observed
- 2026-07-14 04:20:25Z
- Source ID
- Z-2640-2026
- Payload hash
- 0de428ae94c101b9dfc59d9e…
Summary
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing. Status: Ongoing.
Entities
Frequently asked questions
- Has Accuray Incorporated had other FDA enforcement actions?
- Yes. Argus HQ has recorded 2 total FDA actions tied to Accuray Incorporated: 0 warning letters, 2 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2026-07-08; Argus observed and indexed it on 2026-07-14.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2026). Accuray Incorporated — Class II recall: Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator…. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-accuray-incorporated-z-2640-2026
MLA
"Accuray Incorporated — Class II recall: Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator…." Argus HQ Research, Digital Empire LLC, 2026, argushq.ai/events/fda-recall-accuray-incorporated-z-2640-2026.
Chicago
Argus HQ Research. "Accuray Incorporated — Class II recall: Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator…." Digital Empire LLC. Accessed July 14, 2026. https://argushq.ai/events/fda-recall-accuray-incorporated-z-2640-2026.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_accuray_incorporated_z_2640_2026_2026,
title = {Accuray Incorporated — Class II recall: Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator…},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/events/fda-recall-accuray-incorporated-z-2640-2026},
note = {Accessed: July 14, 2026}
}Informational only. Not legal, financial, or medical advice.