Skip to main content

fda_recall · device_recall · critical

Baxter Healthcare Corporation — Class I recall: Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS

Event date
2025-08-20 00:00:00Z
Observed
2026-07-16 23:20:09Z
Source ID
Z-2173-2025
Payload hash
971379dd0831f15de2297723

Summary

Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The level of underinfusion is variable based on the… Status: Ongoing.

Frequently asked questions

Has Baxter Healthcare Corporation had other FDA enforcement actions?
Yes. Argus HQ has recorded 208 total FDA actions tied to Baxter Healthcare Corporation: 0 warning letters, 208 recalls, 0 approval records, and 0 Form 483 inspection citations.
What kind of FDA record is this?
This is an Argus HQ record of a FDA Recall, tagged critical severity by Argus HQ. FDA published the underlying record on 2025-08-20; Argus observed and indexed it on 2026-07-16.
Where can I verify this record directly with FDA?
The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.

Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder

Cite this record

Reusing this data in reporting or research? Here’s a ready-made citation.

APA
Argus HQ Research (2025). Baxter Healthcare Corporation — Class I recall: Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-baxter-healthcare-corporation-z-2173-2025
MLA
"Baxter Healthcare Corporation — Class I recall: Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS." Argus HQ Research, Digital Empire LLC, 2025, argushq.ai/events/fda-recall-baxter-healthcare-corporation-z-2173-2025.
Chicago
Argus HQ Research. "Baxter Healthcare Corporation — Class I recall: Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-baxter-healthcare-corporation-z-2173-2025.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_baxter_healthcare_corporation_z_2173_2025_2025,
  title = {Baxter Healthcare Corporation — Class I recall: Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS},
  author = {{Argus HQ Research}},
  year = {2025},
  publisher = {Digital Empire LLC},
  url = {https://argushq.ai/events/fda-recall-baxter-healthcare-corporation-z-2173-2025},
  note = {Accessed: July 17, 2026}
}

Informational only. Not legal, financial, or medical advice.