fda_recall · device_recall · high
Becton Dickinson & Co. — Class II recall: BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin…
- Event date
- 2024-12-11 00:00:00Z
- Observed
- 2026-07-16 23:20:09Z
- Source ID
- Z-0621-2025
- Payload hash
- 6c06f920ec47266b1ce58f44…
Summary
BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility Status: Ongoing.
Entities
Frequently asked questions
- Has Becton Dickinson & Company had other FDA enforcement actions?
- Yes. Argus HQ has recorded 82 total FDA actions tied to Becton Dickinson & Company: 0 warning letters, 82 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2024-12-11; Argus observed and indexed it on 2026-07-16.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2024). Becton Dickinson & Co. — Class II recall: BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin…. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-becton-dickinson-and-co-z-0621-2025
MLA
"Becton Dickinson & Co. — Class II recall: BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin…." Argus HQ Research, Digital Empire LLC, 2024, argushq.ai/events/fda-recall-becton-dickinson-and-co-z-0621-2025.
Chicago
Argus HQ Research. "Becton Dickinson & Co. — Class II recall: BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin…." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-becton-dickinson-and-co-z-0621-2025.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_becton_dickinson_and_co_z_0621_2025_2024,
title = {Becton Dickinson & Co. — Class II recall: BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin…},
author = {{Argus HQ Research}},
year = {2024},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/events/fda-recall-becton-dickinson-and-co-z-0621-2025},
note = {Accessed: July 17, 2026}
}Informational only. Not legal, financial, or medical advice.