Skip to main content

fda_recall · device_recall · high

Becton, Dickinson and Company, BD Bio Sciences — Class II recall: BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSD…

Event date
2024-10-09 00:00:00Z
Observed
2026-07-16 23:20:09Z
Source ID
Z-0009-2025
Payload hash
94b9cfbf1bdbc6a0c7477133

Summary

Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute coun… Status: Ongoing.

Frequently asked questions

Has Becton, Dickinson and Company, BD Bio Sciences had other FDA enforcement actions?
Yes. Argus HQ has recorded 10 total FDA actions tied to Becton, Dickinson and Company, BD Bio Sciences: 0 warning letters, 10 recalls, 0 approval records, and 0 Form 483 inspection citations.
What kind of FDA record is this?
This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2024-10-09; Argus observed and indexed it on 2026-07-16.
Where can I verify this record directly with FDA?
The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.

Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder

Cite this record

Reusing this data in reporting or research? Here’s a ready-made citation.

APA
Argus HQ Research (2024). Becton, Dickinson and Company, BD Bio Sciences — Class II recall: BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSD…. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-becton-dickinson-and-company-bd-bio-sciences-z-0009-2025
MLA
"Becton, Dickinson and Company, BD Bio Sciences — Class II recall: BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSD…." Argus HQ Research, Digital Empire LLC, 2024, argushq.ai/events/fda-recall-becton-dickinson-and-company-bd-bio-sciences-z-0009-2025.
Chicago
Argus HQ Research. "Becton, Dickinson and Company, BD Bio Sciences — Class II recall: BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSD…." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-becton-dickinson-and-company-bd-bio-sciences-z-0009-2025.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_becton_dickinson_and_company_bd_bio_sciences_z_0009_2025_2024,
  title = {Becton, Dickinson and Company, BD Bio Sciences — Class II recall: BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSD…},
  author = {{Argus HQ Research}},
  year = {2024},
  publisher = {Digital Empire LLC},
  url = {https://argushq.ai/events/fda-recall-becton-dickinson-and-company-bd-bio-sciences-z-0009-2025},
  note = {Accessed: July 17, 2026}
}

Informational only. Not legal, financial, or medical advice.