fda_recall · device_recall · high
Becton, Dickinson and Company, BD Bio Sciences — Class II recall: BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSD…
- Event date
- 2024-10-09 00:00:00Z
- Observed
- 2026-07-16 23:20:09Z
- Source ID
- Z-0009-2025
- Payload hash
- 94b9cfbf1bdbc6a0c7477133…
Summary
Sample preparation system with software may cause: 1) Multidispense feature enabled with 2-tube assay: Specimen volume dispensed into tube1 at least 3% higher than volume in tube2, or 2) Absolute counts calculated using Trucount Tubes: increased volume in tube1 may result in 20% higher absolute coun… Status: Ongoing.
Frequently asked questions
- Has Becton, Dickinson and Company, BD Bio Sciences had other FDA enforcement actions?
- Yes. Argus HQ has recorded 10 total FDA actions tied to Becton, Dickinson and Company, BD Bio Sciences: 0 warning letters, 10 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2024-10-09; Argus observed and indexed it on 2026-07-16.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2024). Becton, Dickinson and Company, BD Bio Sciences — Class II recall: BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSD…. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-becton-dickinson-and-company-bd-bio-sciences-z-0009-2025
MLA
"Becton, Dickinson and Company, BD Bio Sciences — Class II recall: BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSD…." Argus HQ Research, Digital Empire LLC, 2024, argushq.ai/events/fda-recall-becton-dickinson-and-company-bd-bio-sciences-z-0009-2025.
Chicago
Argus HQ Research. "Becton, Dickinson and Company, BD Bio Sciences — Class II recall: BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSD…." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-becton-dickinson-and-company-bd-bio-sciences-z-0009-2025.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_becton_dickinson_and_company_bd_bio_sciences_z_0009_2025_2024,
title = {Becton, Dickinson and Company, BD Bio Sciences — Class II recall: BD FACSDuet Premium Sample Preparation System, REF: 666340, 666339; and BD FACSD…},
author = {{Argus HQ Research}},
year = {2024},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/events/fda-recall-becton-dickinson-and-company-bd-bio-sciences-z-0009-2025},
note = {Accessed: July 17, 2026}