fda_recall · device_recall · high
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Allura Xper FD20 OR Table; Model Numbers: (1) 722015, (2) 722023, (3) 722035;…
- Event date
- 2025-10-08 00:00:00Z
- Observed
- 2026-07-16 23:20:10Z
- Source ID
- Z-0008-2026
- Payload hash
- 8d47d3391f2f0b7010b4ccaf…
Summary
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted. Status: Ongoing.
Frequently asked questions
- Has PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. had other FDA enforcement actions?
- Yes. Argus HQ has recorded 199 total FDA actions tied to PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: 0 warning letters, 199 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2025-10-08; Argus observed and indexed it on 2026-07-16.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2025). PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Allura Xper FD20 OR Table; Model Numbers: (1) 722015, (2) 722023, (3) 722035;…. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-0008-2026
MLA
"PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Allura Xper FD20 OR Table; Model Numbers: (1) 722015, (2) 722023, (3) 722035;…." Argus HQ Research, Digital Empire LLC, 2025, argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-0008-2026.
Chicago
Argus HQ Research. "PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Allura Xper FD20 OR Table; Model Numbers: (1) 722015, (2) 722023, (3) 722035;…." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-0008-2026.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_philips_medical_systems_nederland_b_v_z_0008_2026_2025,
title = {PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Allura Xper FD20 OR Table; Model Numbers: (1) 722015, (2) 722023, (3) 722035;…},
author = {{Argus HQ Research}},
year = {2025},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-0008-2026},
note = {Accessed: July 17, 2026}
}Informational only. Not legal, financial, or medical advice.