fda_recall · device_recall · high
Philips North America — Class II recall: SmartPath to dStream for 1.5T¿; Product Number: 782146;
- Event date
- 2025-06-18 00:00:00Z
- Observed
- 2026-07-16 23:20:10Z
- Source ID
- Z-1936-2025
- Payload hash
- e3b05af030f70b92c6ec27e6…
Summary
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyx… Status: Ongoing.
Entities
Frequently asked questions
- Has Philips North America had other FDA enforcement actions?
- Yes. Argus HQ has recorded 355 total FDA actions tied to Philips North America: 0 warning letters, 355 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2025-06-18; Argus observed and indexed it on 2026-07-16.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2025). Philips North America — Class II recall: SmartPath to dStream for 1.5T¿; Product Number: 782146;. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-philips-north-america-z-1936-2025
MLA
"Philips North America — Class II recall: SmartPath to dStream for 1.5T¿; Product Number: 782146;." Argus HQ Research, Digital Empire LLC, 2025, argushq.ai/events/fda-recall-philips-north-america-z-1936-2025.
Chicago
Argus HQ Research. "Philips North America — Class II recall: SmartPath to dStream for 1.5T¿; Product Number: 782146;." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-philips-north-america-z-1936-2025.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_philips_north_america_z_1936_2025_2025,
title = {Philips North America — Class II recall: SmartPath to dStream for 1.5T¿; Product Number: 782146;},
author = {{Argus HQ Research}},
year = {2025},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/events/fda-recall-philips-north-america-z-1936-2025},
note = {Accessed: July 17, 2026}