Wipro GE Healthcare Private Ltd. — Class II recall: GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software v…

Event date
2025-10-29 00:00:00Z
Observed
2026-07-16 23:20:10Z
Source ID
Z-0213-2026
Payload hash
66c2dcf959529fed20c8055d

Summary

Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120. Status: Ongoing.

Frequently asked questions

Has Wipro GE Healthcare Private Ltd. had other FDA enforcement actions?
Yes. Argus HQ has recorded 10 total FDA actions tied to Wipro GE Healthcare Private Ltd.: 0 warning letters, 10 recalls, 0 approval records, and 0 Form 483 inspection citations.
What kind of FDA record is this?
This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2025-10-29; Argus observed and indexed it on 2026-07-16.
Where can I verify this record directly with FDA?
The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.

Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder

Cite this record

Reusing this data in reporting or research? Here’s a ready-made citation.

APA
Argus HQ Research (2025). Wipro GE Healthcare Private Ltd. — Class II recall: GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software v…. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-wipro-ge-healthcare-private-ltd-z-0213-2026
MLA
"Wipro GE Healthcare Private Ltd. — Class II recall: GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software v…." Argus HQ Research, Digital Empire LLC, 2025, argushq.ai/events/fda-recall-wipro-ge-healthcare-private-ltd-z-0213-2026.
Chicago
Argus HQ Research. "Wipro GE Healthcare Private Ltd. — Class II recall: GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software v…." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-wipro-ge-healthcare-private-ltd-z-0213-2026.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_wipro_ge_healthcare_private_ltd_z_0213_2026_2025,
  title = {Wipro GE Healthcare Private Ltd. — Class II recall: GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software v…},
  author = {{Argus HQ Research}},
  year = {2025},
  publisher = {Digital Empire LLC},
  url = {https://argushq.ai/events/fda-recall-wipro-ge-healthcare-private-ltd-z-0213-2026},
  note = {Accessed: July 17, 2026}
}

Informational only. Not legal, financial, or medical advice.