Summary
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look c… Status: Ongoing.
Entities
Frequently asked questions
- Has Wipro GE Healthcare Private Ltd. had other FDA enforcement actions?
- Yes. Argus HQ has recorded 10 total FDA actions tied to Wipro GE Healthcare Private Ltd.: 0 warning letters, 10 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged critical severity by Argus HQ. FDA published the underlying record on 2024-02-21; Argus observed and indexed it on 2026-07-16.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2024). Wipro GE Healthcare Private Ltd. — Class I recall: GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000,…. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-wipro-ge-healthcare-private-ltd-z-0957-2024
MLA
"Wipro GE Healthcare Private Ltd. — Class I recall: GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000,…." Argus HQ Research, Digital Empire LLC, 2024, argushq.ai/events/fda-recall-wipro-ge-healthcare-private-ltd-z-0957-2024.
Chicago
Argus HQ Research. "Wipro GE Healthcare Private Ltd. — Class I recall: GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000,…." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-wipro-ge-healthcare-private-ltd-z-0957-2024.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_wipro_ge_healthcare_private_ltd_z_0957_2024_2024,
title = {Wipro GE Healthcare Private Ltd. — Class I recall: GE Healthcare neonatal incubator, models: a) Care Plus, b) Care Plus 1000,…},
author = {{Argus HQ Research}},
year = {2024},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/events/fda-recall-wipro-ge-healthcare-private-ltd-z-0957-2024},
note = {Accessed: July 17, 2026}
}Informational only. Not legal, financial, or medical advice.