fda_recall · device_recall · high
Abbott — Class II recall: Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning genera…
- Event date
- 2023-11-08 00:00:00Z
- Observed
- 2026-07-16 23:20:09Z
- Source ID
- Z-0209-2024
- Payload hash
- d5b6e2440129fdd3ee7c5a75…
Summary
Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure. Status: Ongoing.
Entities
Frequently asked questions
- Has Abbott had other FDA enforcement actions?
- Yes. Argus HQ has recorded 11 total FDA actions tied to Abbott: 0 warning letters, 11 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2023-11-08; Argus observed and indexed it on 2026-07-16.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2023). Abbott — Class II recall: Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning genera…. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-abbott-z-0209-2024
MLA
"Abbott — Class II recall: Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning genera…." Argus HQ Research, Digital Empire Holdings LLC, 2023, argushq.ai/events/fda-recall-abbott-z-0209-2024.
Chicago
Argus HQ Research. "Abbott — Class II recall: Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning genera…." Digital Empire Holdings LLC. Accessed July 23, 2026. https://argushq.ai/events/fda-recall-abbott-z-0209-2024.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_abbott_z_0209_2024_2023,
title = {Abbott — Class II recall: Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning genera…},
author = {{Argus HQ Research}},
year = {2023},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/events/fda-recall-abbott-z-0209-2024},
note = {Accessed: July 23, 2026}
}Informational only. Not legal, financial, or medical advice.