Skip to main content

fda_recall · device_recall · high

Boston Scientific Corporation — Class II recall: HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm UPN: M00…

Event date
2025-02-05 00:00:00Z
Observed
2026-07-16 23:20:09Z
Source ID
Z-1035-2025
Payload hash
e534f8a3a40ad648b4bd8fc5

Summary

Reported Outer sheath distal black tip detaching from the device can remain around the stent saddle, preventing proper expansion, resulting in prolongation of the procedure to exchange the device for a new one. Status: Ongoing.

Frequently asked questions

Has Boston Scientific Corporation had other FDA enforcement actions?
Yes. Argus HQ has recorded 272 total FDA actions tied to Boston Scientific Corporation: 0 warning letters, 272 recalls, 0 approval records, and 0 Form 483 inspection citations.
What kind of FDA record is this?
This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2025-02-05; Argus observed and indexed it on 2026-07-16.
Where can I verify this record directly with FDA?
The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.

Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder

Cite this record

Reusing this data in reporting or research? Here’s a ready-made citation.

APA
Argus HQ Research (2025). Boston Scientific Corporation — Class II recall: HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm  UPN: M00…. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-boston-scientific-corporation-z-1035-2025
MLA
"Boston Scientific Corporation — Class II recall: HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm  UPN: M00…." Argus HQ Research, Digital Empire Holdings LLC, 2025, argushq.ai/events/fda-recall-boston-scientific-corporation-z-1035-2025.
Chicago
Argus HQ Research. "Boston Scientific Corporation — Class II recall: HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm  UPN: M00…." Digital Empire Holdings LLC. Accessed July 23, 2026. https://argushq.ai/events/fda-recall-boston-scientific-corporation-z-1035-2025.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_boston_scientific_corporation_z_1035_2025_2025,
  title = {Boston Scientific Corporation — Class II recall: HOT AXIOS Stent and Electrocautery- Enhanced Delivery System 15mmx10mm  UPN: M00…},
  author = {{Argus HQ Research}},
  year = {2025},
  publisher = {Digital Empire Holdings LLC},
  url = {https://argushq.ai/events/fda-recall-boston-scientific-corporation-z-1035-2025},
  note = {Accessed: July 23, 2026}
}

Informational only. Not legal, financial, or medical advice.