fda_recall · device_recall · high
Boston Scientific Corporation — Class II recall: FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculatu…
- Event date
- 2026-07-29 00:00:00Z
- Observed
- 2026-08-04 04:20:55Z
- Source ID
- Z-2754-2026
- Payload hash
- 06e2858aa5600bf2f3c2deeb…
Summary
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient… Status: Ongoing.
Entities
Frequently asked questions
- Has Boston Scientific Corporation had other FDA enforcement actions?
- Yes. Argus HQ has recorded 282 total FDA actions tied to Boston Scientific Corporation: 0 warning letters, 282 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2026-07-29; Argus observed and indexed it on 2026-08-04.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2026). Boston Scientific Corporation — Class II recall: FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculatu…. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-boston-scientific-corporation-z-2754-2026
MLA
"Boston Scientific Corporation — Class II recall: FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculatu…." Argus HQ Research, Digital Empire Holdings LLC, 2026, argushq.ai/events/fda-recall-boston-scientific-corporation-z-2754-2026.
Chicago
Argus HQ Research. "Boston Scientific Corporation — Class II recall: FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculatu…." Digital Empire Holdings LLC. Accessed August 4, 2026. https://argushq.ai/events/fda-recall-boston-scientific-corporation-z-2754-2026.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_boston_scientific_corporation_z_2754_2026_2026,
title = {Boston Scientific Corporation — Class II recall: FARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculatu…},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/events/fda-recall-boston-scientific-corporation-z-2754-2026},
note = {Accessed: August 4, 2026}
}Informational only. Not legal, financial, or medical advice.