Daavlin Distributing Company — Class II recall: Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
- Event date
- 2026-07-22 00:00:00Z
- Observed
- 2026-07-28 04:20:57Z
- Source ID
- Z-2743-2026
- Payload hash
- 8083ccb6ae3aba74314318d8…
Summary
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000). Status: Ongoing.
Entities
Frequently asked questions
- Has Daavlin Distributing Company had other FDA enforcement actions?
- Yes. Argus HQ has recorded 4 total FDA actions tied to Daavlin Distributing Company: 0 warning letters, 4 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2026-07-22; Argus observed and indexed it on 2026-07-28.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2026). Daavlin Distributing Company — Class II recall: Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-daavlin-distributing-company-z-2743-2026
MLA
"Daavlin Distributing Company — Class II recall: Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders." Argus HQ Research, Digital Empire Holdings LLC, 2026, argushq.ai/events/fda-recall-daavlin-distributing-company-z-2743-2026.
Chicago
Argus HQ Research. "Daavlin Distributing Company — Class II recall: Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders." Digital Empire Holdings LLC. Accessed July 28, 2026. https://argushq.ai/events/fda-recall-daavlin-distributing-company-z-2743-2026.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_daavlin_distributing_company_z_2743_2026_2026,
title = {Daavlin Distributing Company — Class II recall: Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/events/fda-recall-daavlin-distributing-company-z-2743-2026},
note = {Accessed: July 28, 2026}
}Informational only. Not legal, financial, or medical advice.