fda_recall · device_recall · high
GE Medical Systems Information Technologies Inc — Class II recall: GE HealthCare ApexPro v6.0 eDelivery Software, REF #5700026, Product Code SUV; u…
- Event date
- 2026-06-03 00:00:00Z
- Observed
- 2026-07-09 22:52:23Z
- Source ID
- Z-2247-2026
- Payload hash
- 24bfff50c6fe496220d8901b…
Summary
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Cen… Status: Ongoing.
Informational only. Not legal, financial, or medical advice.