fda_recall · device_recall · high

GE Medical Systems Information Technologies Inc — Class II recall: GE HealthCare ApexPro v6.0 eDelivery Software, REF #5700026, Product Code SUV; u…

Event date
2026-06-03 00:00:00Z
Observed
2026-07-09 22:52:23Z
Source ID
Z-2247-2026
Payload hash
24bfff50c6fe496220d8901b

Summary

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Cen… Status: Ongoing.