fda_recall · device_recall · high
ICU Medical, Inc. — Class II recall: Plum Solo Precision IV Pump, 40001-0401
- Event date
- 2026-06-17 00:00:00Z
- Observed
- 2026-07-09 22:52:16Z
- Source ID
- Z-2311-2026
- Payload hash
- fb8a55e9818b3c57c9182ea4…
Summary
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery. Status: Ongoing.