fda_recall · device_recall · high
Intuitive Surgical, Inc. — Class II recall: Da Vinci ASSY, PSS, SP1098, Part Number: 380601
- Event date
- 2026-07-01 00:00:00Z
- Observed
- 2026-07-09 22:51:53Z
- Source ID
- Z-2521-2026
- Payload hash
- 4b06c4fa833c6d4779902f97…
Summary
Operative components of surgical systems potentially distributed with E-brake retainer pin, so in event of a structural column failure, the E-brake pin would prevent E-brake engagement and may result in uncontrolled motion. In addition, if it moves/backs out it may cause restricted column range of m… Status: Ongoing.
Entities