fda_recall · device_recall · high
Medtronic, Inc. — Class II recall: Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the follow…
- Event date
- 2025-07-30 00:00:00Z
- Observed
- 2026-07-16 23:20:07Z
- Source ID
- Z-2169-2025
- Payload hash
- b48b79697a3febd4de66ff40…
Summary
In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery. Status: Ongoing.
Entities
Frequently asked questions
- Has Medtronic, Inc. had other FDA enforcement actions?
- Yes. Argus HQ has recorded 40 total FDA actions tied to Medtronic, Inc.: 0 warning letters, 40 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2025-07-30; Argus observed and indexed it on 2026-07-16.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2025). Medtronic, Inc. — Class II recall: Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the follow…. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-medtronic-inc-z-2169-2025
MLA
"Medtronic, Inc. — Class II recall: Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the follow…." Argus HQ Research, Digital Empire Holdings LLC, 2025, argushq.ai/events/fda-recall-medtronic-inc-z-2169-2025.
Chicago
Argus HQ Research. "Medtronic, Inc. — Class II recall: Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the follow…." Digital Empire Holdings LLC. Accessed July 23, 2026. https://argushq.ai/events/fda-recall-medtronic-inc-z-2169-2025.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_medtronic_inc_z_2169_2025_2025,
title = {Medtronic, Inc. — Class II recall: Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the follow…},
author = {{Argus HQ Research}},
year = {2025},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/events/fda-recall-medtronic-inc-z-2169-2025},
note = {Accessed: July 23, 2026}
}Informational only. Not legal, financial, or medical advice.