Skip to main content

fda_recall · device_recall · critical

Medtronic MiniMed, Inc. — Class I recall: MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755…

Event date
2024-10-09 00:00:00Z
Observed
2026-07-16 23:20:08Z
Source ID
Z-0001-2025
Payload hash
b4d9fef37ff6951176cee51c

Summary

Insulin pumps that have been dropped, bumped, or experienced physical impact may have damage to internal electrical components, which may cause reduced pump battery life. Since the defect reduces the battery life overall, it shortens the time between low battery alarms and increases the frequency wi… Status: Ongoing.

Frequently asked questions

Has Medtronic MiniMed, Inc. had other FDA enforcement actions?
Yes. Argus HQ has recorded 38 total FDA actions tied to Medtronic MiniMed, Inc.: 0 warning letters, 38 recalls, 0 approval records, and 0 Form 483 inspection citations.
What kind of FDA record is this?
This is an Argus HQ record of a FDA Recall, tagged critical severity by Argus HQ. FDA published the underlying record on 2024-10-09; Argus observed and indexed it on 2026-07-16.
Where can I verify this record directly with FDA?
The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.

Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder

Cite this record

Reusing this data in reporting or research? Here’s a ready-made citation.

APA
Argus HQ Research (2024). Medtronic MiniMed, Inc. — Class I recall: MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755…. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-medtronic-minimed-inc-z-0001-2025
MLA
"Medtronic MiniMed, Inc. — Class I recall: MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755…." Argus HQ Research, Digital Empire Holdings LLC, 2024, argushq.ai/events/fda-recall-medtronic-minimed-inc-z-0001-2025.
Chicago
Argus HQ Research. "Medtronic MiniMed, Inc. — Class I recall: MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755…." Digital Empire Holdings LLC. Accessed July 23, 2026. https://argushq.ai/events/fda-recall-medtronic-minimed-inc-z-0001-2025.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_medtronic_minimed_inc_z_0001_2025_2024,
  title = {Medtronic MiniMed, Inc. — Class I recall: MiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755…},
  author = {{Argus HQ Research}},
  year = {2024},
  publisher = {Digital Empire Holdings LLC},
  url = {https://argushq.ai/events/fda-recall-medtronic-minimed-inc-z-0001-2025},
  note = {Accessed: July 23, 2026}
}

Informational only. Not legal, financial, or medical advice.