fda_recall · device_recall · high
Medtronic MiniMed, Inc. — Class II recall: MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Ve…
- Event date
- 2026-02-11 00:00:00Z
- Observed
- 2026-07-16 23:20:07Z
- Source ID
- Z-1238-2026
- Payload hash
- e87c5ea312f0075fabf90741…
Summary
A software timing and processor communication issue in MiniMed" 780G software version 6.60 may trigger Pump Error 43 and/or Pump Error 41, resulting in suspension of insulin delivery. Status: Ongoing.
Entities
Frequently asked questions
- Has Medtronic MiniMed, Inc. had other FDA enforcement actions?
- Yes. Argus HQ has recorded 38 total FDA actions tied to Medtronic MiniMed, Inc.: 0 warning letters, 38 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2026-02-11; Argus observed and indexed it on 2026-07-16.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2026). Medtronic MiniMed, Inc. — Class II recall: MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Ve…. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-medtronic-minimed-inc-z-1238-2026
MLA
"Medtronic MiniMed, Inc. — Class II recall: MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Ve…." Argus HQ Research, Digital Empire Holdings LLC, 2026, argushq.ai/events/fda-recall-medtronic-minimed-inc-z-1238-2026.
Chicago
Argus HQ Research. "Medtronic MiniMed, Inc. — Class II recall: MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Ve…." Digital Empire Holdings LLC. Accessed July 23, 2026. https://argushq.ai/events/fda-recall-medtronic-minimed-inc-z-1238-2026.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_medtronic_minimed_inc_z_1238_2026_2026,
title = {Medtronic MiniMed, Inc. — Class II recall: MiniMed 780G Insulin Pump Catalog Numbers: MMT-1884 MMT-1886 With Software Ve…},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/events/fda-recall-medtronic-minimed-inc-z-1238-2026},
note = {Accessed: July 23, 2026}
}Informational only. Not legal, financial, or medical advice.