Skip to main content

fda_recall · device_recall · critical

Medtronic Navigation, Inc. — Class I recall: StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit),…

Event date
2023-06-14 00:00:00Z
Observed
2026-07-16 23:20:09Z
Source ID
Z-1630-2023
Payload hash
65c93666b8e5f93786ed0e6a

Summary

During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue… Status: Ongoing.

Frequently asked questions

Has Medtronic Navigation, Inc. had other FDA enforcement actions?
Yes. Argus HQ has recorded 10 total FDA actions tied to Medtronic Navigation, Inc.: 0 warning letters, 10 recalls, 0 approval records, and 0 Form 483 inspection citations.
What kind of FDA record is this?
This is an Argus HQ record of a FDA Recall, tagged critical severity by Argus HQ. FDA published the underlying record on 2023-06-14; Argus observed and indexed it on 2026-07-16.
Where can I verify this record directly with FDA?
The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.

Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder

Cite this record

Reusing this data in reporting or research? Here’s a ready-made citation.

APA
Argus HQ Research (2023). Medtronic Navigation, Inc. — Class I recall: StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit),…. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-medtronic-navigation-inc-z-1630-2023
MLA
"Medtronic Navigation, Inc. — Class I recall: StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit),…." Argus HQ Research, Digital Empire Holdings LLC, 2023, argushq.ai/events/fda-recall-medtronic-navigation-inc-z-1630-2023.
Chicago
Argus HQ Research. "Medtronic Navigation, Inc. — Class I recall: StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit),…." Digital Empire Holdings LLC. Accessed July 23, 2026. https://argushq.ai/events/fda-recall-medtronic-navigation-inc-z-1630-2023.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_medtronic_navigation_inc_z_1630_2023_2023,
  title = {Medtronic Navigation, Inc. — Class I recall: StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit),…},
  author = {{Argus HQ Research}},
  year = {2023},
  publisher = {Digital Empire Holdings LLC},
  url = {https://argushq.ai/events/fda-recall-medtronic-navigation-inc-z-1630-2023},
  note = {Accessed: July 23, 2026}
}