fda_recall · device_recall · high
Medtronic Perfusion Systems — Class II recall: Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product…
- Event date
- 2026-03-04 00:00:00Z
- Observed
- 2026-07-16 23:20:07Z
- Source ID
- Z-1473-2026
- Payload hash
- 643d6f73f76895517cb13d9a…
Summary
During the manufacturing, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design. Status: Ongoing.
Entities
Frequently asked questions
- Has Medtronic Perfusion Systems had other FDA enforcement actions?
- Yes. Argus HQ has recorded 46 total FDA actions tied to Medtronic Perfusion Systems: 0 warning letters, 46 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2026-03-04; Argus observed and indexed it on 2026-07-16.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2026). Medtronic Perfusion Systems — Class II recall: Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product…. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-medtronic-perfusion-systems-z-1473-2026
MLA
"Medtronic Perfusion Systems — Class II recall: Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product…." Argus HQ Research, Digital Empire Holdings LLC, 2026, argushq.ai/events/fda-recall-medtronic-perfusion-systems-z-1473-2026.
Chicago
Argus HQ Research. "Medtronic Perfusion Systems — Class II recall: Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product…." Digital Empire Holdings LLC. Accessed July 23, 2026. https://argushq.ai/events/fda-recall-medtronic-perfusion-systems-z-1473-2026.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_medtronic_perfusion_systems_z_1473_2026_2026,
title = {Medtronic Perfusion Systems — Class II recall: Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product…},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/events/fda-recall-medtronic-perfusion-systems-z-1473-2026},
note = {Accessed: July 23, 2026}
}Informational only. Not legal, financial, or medical advice.