fda_recall · device_recall · critical
Nova Biomedical Corporation — Class I recall: StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.
- Event date
- 2026-09-23 00:00:00Z
- Observed
- 2026-09-29 04:20:32Z
- Source ID
- Z-3128-2026
- Payload hash
- e383a4c0e2f571d5d1edd77f…
Summary
A software synchronization/timing issue in the barcode scanning and patient data processing component that causes the device to intermittently retain a previously scanned barcode including patient identification (ID) barcodes and quality control (QC) solution ID barcodes when a new barcode is sc… Status: Ongoing.
Entities
Frequently asked questions
- Has Nova Biomedical Corporation had other FDA enforcement actions?
- Yes. Argus HQ has recorded 5 total FDA actions tied to Nova Biomedical Corporation: 0 warning letters, 5 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged critical severity by Argus HQ. FDA published the underlying record on 2026-09-23; Argus observed and indexed it on 2026-09-29.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2026). Nova Biomedical Corporation — Class I recall: StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-nova-biomedical-corporation-z-3128-2026
MLA
"Nova Biomedical Corporation — Class I recall: StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.." Argus HQ Research, Digital Empire Holdings LLC, 2026, argushq.ai/events/fda-recall-nova-biomedical-corporation-z-3128-2026.
Chicago
Argus HQ Research. "Nova Biomedical Corporation — Class I recall: StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.." Digital Empire Holdings LLC. Accessed September 29, 2026. https://argushq.ai/events/fda-recall-nova-biomedical-corporation-z-3128-2026.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_nova_biomedical_corporation_z_3128_2026_2026,
title = {Nova Biomedical Corporation — Class I recall: StatStrip Glucose Hospital Meter System. Model No: 63685. Glucose test system.},