PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Intera 1.5T Master/Nova, Model Number: 781106;

Event date
2025-05-21 00:00:00Z
Observed
2026-07-16 23:20:10Z
Source ID
Z-1776-2025
Payload hash
1281e3857239059f7efae1bb

Summary

The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury. Status: Ongoing.

Frequently asked questions

Has PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. had other FDA enforcement actions?
Yes. Argus HQ has recorded 199 total FDA actions tied to PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: 0 warning letters, 199 recalls, 0 approval records, and 0 Form 483 inspection citations.
What kind of FDA record is this?
This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2025-05-21; Argus observed and indexed it on 2026-07-16.
Where can I verify this record directly with FDA?
The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.

Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder

Cite this record

Reusing this data in reporting or research? Here’s a ready-made citation.

APA
Argus HQ Research (2025). PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Intera 1.5T Master/Nova, Model Number: 781106;. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1776-2025
MLA
"PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Intera 1.5T Master/Nova, Model Number: 781106;." Argus HQ Research, Digital Empire LLC, 2025, argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1776-2025.
Chicago
Argus HQ Research. "PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Intera 1.5T Master/Nova, Model Number: 781106;." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1776-2025.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_philips_medical_systems_nederland_b_v_z_1776_2025_2025,
  title = {PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Intera 1.5T Master/Nova, Model Number: 781106;},
  author = {{Argus HQ Research}},
  year = {2025},
  publisher = {Digital Empire LLC},
  url = {https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1776-2025},
  note = {Accessed: July 17, 2026}
}

Informational only. Not legal, financial, or medical advice.