Skip to main content

fda_recall · device_recall · high

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722…

Event date
2024-06-05 00:00:00Z
Observed
2026-07-16 23:20:11Z
Source ID
Z-1895-2024
Payload hash
034a9420332aefb472abca40

Summary

When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table. Status: Ongoing.

Frequently asked questions

Has PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. had other FDA enforcement actions?
Yes. Argus HQ has recorded 199 total FDA actions tied to PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: 0 warning letters, 199 recalls, 0 approval records, and 0 Form 483 inspection citations.
What kind of FDA record is this?
This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2024-06-05; Argus observed and indexed it on 2026-07-16.
Where can I verify this record directly with FDA?
The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.

Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder

Cite this record

Reusing this data in reporting or research? Here’s a ready-made citation.

APA
Argus HQ Research (2024). PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Philips Allura Xper FD20 system with an ADN7NT patient table.  Model Number: 722…. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1895-2024
MLA
"PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Philips Allura Xper FD20 system with an ADN7NT patient table.  Model Number: 722…." Argus HQ Research, Digital Empire LLC, 2024, argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1895-2024.
Chicago
Argus HQ Research. "PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Philips Allura Xper FD20 system with an ADN7NT patient table.  Model Number: 722…." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1895-2024.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_philips_medical_systems_nederland_b_v_z_1895_2024_2024,
  title = {PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Philips Allura Xper FD20 system with an ADN7NT patient table.  Model Number: 722…},
  author = {{Argus HQ Research}},
  year = {2024},
  publisher = {Digital Empire LLC},
  url = {https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1895-2024},
  note = {Accessed: July 17, 2026}
}

Informational only. Not legal, financial, or medical advice.