fda_recall · device_recall · high
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722…
- Event date
- 2024-06-05 00:00:00Z
- Observed
- 2026-07-16 23:20:11Z
- Source ID
- Z-1895-2024
- Payload hash
- 034a9420332aefb472abca40…
Summary
When Software upgrade 8.1.100 is installed on an Allura Xper with an AD7NT (non tilt) table, the pivot brake does not engage. This issue may cause unexpected movements of the table. Status: Ongoing.
Frequently asked questions
- Has PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. had other FDA enforcement actions?
- Yes. Argus HQ has recorded 199 total FDA actions tied to PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: 0 warning letters, 199 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2024-06-05; Argus observed and indexed it on 2026-07-16.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2024). PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722…. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1895-2024
MLA
"PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722…." Argus HQ Research, Digital Empire LLC, 2024, argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1895-2024.
Chicago
Argus HQ Research. "PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722…." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1895-2024.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_philips_medical_systems_nederland_b_v_z_1895_2024_2024,
title = {PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. — Class II recall: Philips Allura Xper FD20 system with an ADN7NT patient table. Model Number: 722…},
author = {{Argus HQ Research}},
year = {2024},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/events/fda-recall-philips-medical-systems-nederland-b-v-z-1895-2024},
note = {Accessed: July 17, 2026}
}Informational only. Not legal, financial, or medical advice.