Skip to main content

fda_recall · device_recall · high

Philips North America — Class II recall: Patient Information Center iX, Product Code 866386, and Patient Information Cent…

Event date
2024-10-16 00:00:00Z
Observed
2026-07-16 23:20:11Z
Source ID
Z-0035-2025
Payload hash
789d8a8be98731bf6c958766

Summary

During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize t… Status: Ongoing.

Frequently asked questions

Has Philips North America had other FDA enforcement actions?
Yes. Argus HQ has recorded 355 total FDA actions tied to Philips North America: 0 warning letters, 355 recalls, 0 approval records, and 0 Form 483 inspection citations.
What kind of FDA record is this?
This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2024-10-16; Argus observed and indexed it on 2026-07-16.
Where can I verify this record directly with FDA?
The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.

Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder

Cite this record

Reusing this data in reporting or research? Here’s a ready-made citation.

APA
Argus HQ Research (2024). Philips North America — Class II recall: Patient Information Center iX, Product Code 866386, and Patient Information Cent…. Digital Empire LLC. Retrieved from https://argushq.ai/events/fda-recall-philips-north-america-z-0035-2025
MLA
"Philips North America — Class II recall: Patient Information Center iX, Product Code 866386, and Patient Information Cent…." Argus HQ Research, Digital Empire LLC, 2024, argushq.ai/events/fda-recall-philips-north-america-z-0035-2025.
Chicago
Argus HQ Research. "Philips North America — Class II recall: Patient Information Center iX, Product Code 866386, and Patient Information Cent…." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/events/fda-recall-philips-north-america-z-0035-2025.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_philips_north_america_z_0035_2025_2024,
  title = {Philips North America — Class II recall: Patient Information Center iX, Product Code 866386, and Patient Information Cent…},
  author = {{Argus HQ Research}},
  year = {2024},
  publisher = {Digital Empire LLC},
  url = {https://argushq.ai/events/fda-recall-philips-north-america-z-0035-2025},
  note = {Accessed: July 17, 2026}
}

Informational only. Not legal, financial, or medical advice.