fda_recall · device_recall · high
Philips North America — Class II recall: MR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Softw…
- Event date
- 2026-10-07 00:00:00Z
- Observed
- 2026-10-13 04:20:26Z
- Source ID
- Z-3313-2026
- Payload hash
- f7bad1dde1ea9ef3f86c3658…
Summary
When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathway… Status: Ongoing.
Entities
Frequently asked questions
- Has Philips North America had other FDA enforcement actions?
- Yes. Argus HQ has recorded 369 total FDA actions tied to Philips North America: 0 warning letters, 369 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2026-10-07; Argus observed and indexed it on 2026-10-13.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2026). Philips North America — Class II recall: MR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Softw…. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-philips-north-america-z-3313-2026
MLA
"Philips North America — Class II recall: MR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Softw…." Argus HQ Research, Digital Empire Holdings LLC, 2026, argushq.ai/events/fda-recall-philips-north-america-z-3313-2026.
Chicago
Argus HQ Research. "Philips North America — Class II recall: MR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Softw…." Digital Empire Holdings LLC. Accessed October 13, 2026. https://argushq.ai/events/fda-recall-philips-north-america-z-3313-2026.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_philips_north_america_z_3313_2026_2026,
title = {Philips North America — Class II recall: MR 7700; Model number (REF): 1. 782120, 2. 782153, 3. 782155, 4. 782199; Softw…},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/events/fda-recall-philips-north-america-z-3313-2026},
note = {Accessed: October 13, 2026}
}Informational only. Not legal, financial, or medical advice.