fda_recall · device_recall · high
Roche Diagnostics Operations, Inc. — Class II recall: cobas pro integrated solutions with cobas c 503 analytical units: cobas pro s…
- Event date
- 2026-03-11 00:00:00Z
- Observed
- 2026-07-16 23:20:07Z
- Source ID
- Z-1476-2026
- Payload hash
- 09fe5d37d507b4a207e1f93c…
Summary
Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as io… Status: Ongoing.
Frequently asked questions
- Has Roche Diagnostics Operations, Inc. had other FDA enforcement actions?
- Yes. Argus HQ has recorded 6 total FDA actions tied to Roche Diagnostics Operations, Inc.: 0 warning letters, 6 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2026-03-11; Argus observed and indexed it on 2026-07-16.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2026). Roche Diagnostics Operations, Inc. — Class II recall: cobas pro integrated solutions with cobas c 503 analytical units: cobas pro s…. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-roche-diagnostics-operations-inc-z-1476-2026
MLA
"Roche Diagnostics Operations, Inc. — Class II recall: cobas pro integrated solutions with cobas c 503 analytical units: cobas pro s…." Argus HQ Research, Digital Empire Holdings LLC, 2026, argushq.ai/events/fda-recall-roche-diagnostics-operations-inc-z-1476-2026.
Chicago
Argus HQ Research. "Roche Diagnostics Operations, Inc. — Class II recall: cobas pro integrated solutions with cobas c 503 analytical units: cobas pro s…." Digital Empire Holdings LLC. Accessed July 23, 2026. https://argushq.ai/events/fda-recall-roche-diagnostics-operations-inc-z-1476-2026.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_roche_diagnostics_operations_inc_z_1476_2026_2026,
title = {Roche Diagnostics Operations, Inc. — Class II recall: cobas pro integrated solutions with cobas c 503 analytical units: cobas pro s…},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/events/fda-recall-roche-diagnostics-operations-inc-z-1476-2026},
note = {Accessed: July 23, 2026}
}Informational only. Not legal, financial, or medical advice.