fda_recall · device_recall · high
Siemens Healthcare Diagnostics, Inc. — Class II recall: Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.
- Event date
- 2026-07-15 00:00:00Z
- Observed
- 2026-07-23 04:20:40Z
- Source ID
- Z-2703-2026
- Payload hash
- 4e420bcd5d39abb240159bc5…
Summary
The potential for intermittent well-to-well bias on the A1c_E assay on the Atellica CH and Atellica CI analyzer which could cause falsely depressed or falsely elevated HbA1c patient results. Status: Ongoing.
Frequently asked questions
- Has Siemens Healthcare Diagnostics, Inc. had other FDA enforcement actions?
- Yes. Argus HQ has recorded 4 total FDA actions tied to Siemens Healthcare Diagnostics, Inc.: 0 warning letters, 4 recalls, 0 approval records, and 0 Form 483 inspection citations.
- What kind of FDA record is this?
- This is an Argus HQ record of a FDA Recall, tagged high severity by Argus HQ. FDA published the underlying record on 2026-07-15; Argus observed and indexed it on 2026-07-23.
- Where can I verify this record directly with FDA?
- The verbatim FDA source record is the source of truth — Argus HQ links directly to it below. Always confirm against fda.gov before acting.
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
APA
Argus HQ Research (2026). Siemens Healthcare Diagnostics, Inc. — Class II recall: Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/events/fda-recall-siemens-healthcare-diagnostics-inc-z-2703-2026
MLA
"Siemens Healthcare Diagnostics, Inc. — Class II recall: Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.." Argus HQ Research, Digital Empire Holdings LLC, 2026, argushq.ai/events/fda-recall-siemens-healthcare-diagnostics-inc-z-2703-2026.
Chicago
Argus HQ Research. "Siemens Healthcare Diagnostics, Inc. — Class II recall: Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.." Digital Empire Holdings LLC. Accessed July 23, 2026. https://argushq.ai/events/fda-recall-siemens-healthcare-diagnostics-inc-z-2703-2026.
BibTeX
@misc{argushq_argushq_ai_events_fda_recall_siemens_healthcare_diagnostics_inc_z_2703_2026_2026,
title = {Siemens Healthcare Diagnostics, Inc. — Class II recall: Atellica CH Enzymatic Hemoglobin A1c (A1c_E). Material Number: 11097536.},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/events/fda-recall-siemens-healthcare-diagnostics-inc-z-2703-2026},
note = {Accessed: July 23, 2026}
}Informational only. Not legal, financial, or medical advice.