Key facts
| Recalling firm | Abbott GmbH |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | Alinity i Anti-HCV Reagent Kit, List Number 08P0521 |
| Classification | Class II |
| Recall number | Z-1480-2024 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05). |
What consumers should do
Facilities, clinicians, and consumers holding an affected unit or lot of Alinity i Anti-HCV Reagent Kit, List Number 08P0521 should stop using it for the purpose described in the recall notice and hold remaining stock pending instructions from Abbott GmbH. FDA recall records typically direct the recalling firm to notify known accounts directly with a recall letter describing the specific corrective action -- replacement, field correction, refund, or return. Anyone who received the product through a distributor rather than directly from Abbott GmbH should confirm with their supplier whether their specific lot or unit is on the affected list, since recall notices are lot- or unit-specific and not every unit of a product line is necessarily included. Questions about a specific unit's status should go to Abbott GmbH directly or to the distributor that supplied it; Argus HQ is a public-record aggregator, not the recalling firm, and cannot confirm individual unit status.
What this classification means
A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote. FDA still requires the recalling firm to notify affected accounts and submit status reports on the correction, though the monitoring intensity is lower than for a Class I recall. FDA's Center for Devices and Radiological Health (for devices) or Center for Drug Evaluation and Research (for drugs) reviews the classification and can require additional corrective action if the initial response is judged insufficient. Because this record is filed as Class II, professionals and consumers handling an affected lot should treat FDA's own classification -- not this page's summary of it -- as the authoritative risk signal. A Class I recall is FDA’s most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case -- it reflects FDA’s assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field.
Abbott GmbH’s FDA history
Argus HQ has recorded 2 total FDA actions tied to Abbott GmbH: 0 warning letters, 2 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- Where can I verify this recall directly with FDA?
- The verbatim FDA record is the source of truth for every recall Argus HQ tracks — always confirm against fda.gov before acting.
- Is Alinity i Anti-HCV Reagent Kit, List Number 08P0521 recalled?
- Yes. FDA lists a Class II recall by Abbott GmbH for Alinity i Anti-HCV Reagent Kit, List Number 08P0521, reported 20240410, with status "Ongoing." FDA recall number: Z-1480-2024.
- What is the recall class for Abbott GmbH's recall?
- FDA classified this recall as Class II. A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
- Why was Alinity i Anti-HCV Reagent Kit, List Number 08P0521 recalled?
- FDA's recall record states the reason as: Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Abbott GmbHCompiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2024). FDA Class II Recall: Alinity i Anti-HCV Reagent Kit, List Number 08P0521 by Abbott GmbH — Recall Details. Digital Empire LLC. Retrieved from https://argushq.ai/recall/abbott-z-1480-2024-2024-04-10
"FDA Class II Recall: Alinity i Anti-HCV Reagent Kit, List Number 08P0521 by Abbott GmbH — Recall Details." Argus HQ Research, Digital Empire LLC, 2024, argushq.ai/recall/abbott-z-1480-2024-2024-04-10.
Argus HQ Research. "FDA Class II Recall: Alinity i Anti-HCV Reagent Kit, List Number 08P0521 by Abbott GmbH — Recall Details." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/recall/abbott-z-1480-2024-2024-04-10.
@misc{argushq_argushq_ai_recall_abbott_z_1480_2024_2024_04_10_2024,
title = {FDA Class II Recall: Alinity i Anti-HCV Reagent Kit, List Number 08P0521 by Abbott GmbH — Recall Details},
author = {{Argus HQ Research}},
year = {2024},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/recall/abbott-z-1480-2024-2024-04-10},
note = {Accessed: July 17, 2026}
}Source: FDA.gov — Recalls ↗