Key facts
| Recalling firm | Boston Scientific Corporation |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | WATCHMAN TruSeal Access System DBL, OUS, Material Number (UPN) M635TS70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems. |
| Classification | Class I |
| Recall number | Z-2427-2025 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse. |
What this classification means
A Class I recall is FDA’s most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case -- it reflects FDA’s assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field. FDA's Center for Devices and Radiological Health (for devices) or Center for Drug Evaluation and Research (for drugs) reviews the classification and can require additional corrective action if the initial response is judged insufficient. Because this record is filed as Class I, professionals and consumers handling an affected lot should treat FDA's own classification -- not this page's summary of it -- as the authoritative risk signal. A Class I recall is FDA’s most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case -- it reflects FDA’s assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field.
What consumers should do
Facilities, clinicians, and consumers holding an affected unit or lot of WATCHMAN TruSeal Access System DBL, OUS, Material Number (UPN) M635TS70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems. should stop using it for the purpose described in the recall notice and hold remaining stock pending instructions from Boston Scientific Corporation. FDA recall records typically direct the recalling firm to notify known accounts directly with a recall letter describing the specific corrective action -- replacement, field correction, refund, or return. Anyone who received the product through a distributor rather than directly from Boston Scientific Corporation should confirm with their supplier whether their specific lot or unit is on the affected list, since recall notices are lot- or unit-specific and not every unit of a product line is necessarily included. Questions about a specific unit's status should go to Boston Scientific Corporation directly or to the distributor that supplied it; Argus HQ is a public-record aggregator, not the recalling firm, and cannot confirm individual unit status.
Boston Scientific Corporation’s FDA history
Argus HQ has recorded 272 total FDA actions tied to Boston Scientific Corporation: 0 warning letters, 272 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- Is WATCHMAN TruSeal Access System DBL, OUS, Material Number (UP recalled?
- Yes. FDA lists a Class I recall by Boston Scientific Corporation for WATCHMAN TruSeal Access System DBL, OUS, Material Number (UPN) M635TS70020; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems., reported 20250910, with status "Ongoing." FDA recall number: Z-2427-2025.
- What is the recall class for Boston Scientific Corporation's recall?
- FDA classified this recall as Class I. A Class I recall is FDA’s most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death.
- Why was WATCHMAN TruSeal Access System DBL, OUS, Material Number (UP recalled?
- FDA's recall record states the reason as: Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
- Who should I contact about this recall?
- Contact Boston Scientific Corporation directly, or the distributor/pharmacy that supplied the product, to confirm whether your specific lot or unit is affected. Argus HQ aggregates the public FDA recall record and is not the recalling firm.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Boston Scientific Corporation- Recall: Boston Scientific Corporation — Class II recall: CRE Pro Wireguided 8-10mm 240cm →
- Recall: Boston Scientific Corporation — Class II recall: CRE Wireguided 15-18mm 240cm →
- Recall: Boston Scientific Corporation — Class II recall: CRE Wireguided 10-12mm 240cm →
- Recall: Boston Scientific Corporation — Class II recall: CRE Pro Wireguided 6-8mm 240cm →
- Recall: Boston Scientific Corporation — Class II recall: CRE Wireguided 6-8mm 240cm →
- Recall: Boston Scientific Corporation — Class II recall: CRE Wireguided 18-20mm 240cm →
- Recall: Boston Scientific Corporation — Class II recall: CRE Pro Wireguided 15-18mm 240cm →
- Recall: Boston Scientific Corporation — Class II recall: CRE Pro Wireguided 12-15mm 240cm →
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2025). FDA Recall Database: Class I Recall: WATCHMAN TruSeal Access System DBL, OUS, Material Number… by Boston Scientific Corporation. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/recall/boston-scientific-corporation-z-2427-2025-2025-09-10
"FDA Recall Database: Class I Recall: WATCHMAN TruSeal Access System DBL, OUS, Material Number… by Boston Scientific Corporation." Argus HQ Research, Digital Empire Holdings LLC, 2025, argushq.ai/recall/boston-scientific-corporation-z-2427-2025-2025-09-10.
Argus HQ Research. "FDA Recall Database: Class I Recall: WATCHMAN TruSeal Access System DBL, OUS, Material Number… by Boston Scientific Corporation." Digital Empire Holdings LLC. Accessed July 24, 2026. https://argushq.ai/recall/boston-scientific-corporation-z-2427-2025-2025-09-10.
@misc{argushq_argushq_ai_recall_boston_scientific_corporation_z_2427_2025_2025_09_10_2025,
title = {FDA Recall Database: Class I Recall: WATCHMAN TruSeal Access System DBL, OUS, Material Number… by Boston Scientific Corporation},
author = {{Argus HQ Research}},
year = {2025},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/recall/boston-scientific-corporation-z-2427-2025-2025-09-10},
note = {Accessed: July 24, 2026}
}Source: FDA.gov — Recalls ↗