Key facts
| Recalling firm | Boston Scientific Corporation |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS EU, Model Number 7246, All Versions (Current 7.7) |
| Classification | Class II |
| Recall number | Z-3082-2026 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia. |
What this classification means
A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote. FDA still requires the recalling firm to notify affected accounts and submit status reports on the correction, though the monitoring intensity is lower than for a Class I recall. FDA's Center for Devices and Radiological Health (for devices) or Center for Drug Evaluation and Research (for drugs) reviews the classification and can require additional corrective action if the initial response is judged insufficient. Because this record is filed as Class II, professionals and consumers handling an affected lot should treat FDA's own classification -- not this page's summary of it -- as the authoritative risk signal. A Class I recall is FDA’s most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case -- it reflects FDA’s assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field.
What consumers should do
Facilities, clinicians, and consumers holding an affected unit or lot of LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS EU, Model Number 7246, All Versions (Current 7.7) should stop using it for the purpose described in the recall notice and hold remaining stock pending instructions from Boston Scientific Corporation. FDA recall records typically direct the recalling firm to notify known accounts directly with a recall letter describing the specific corrective action -- replacement, field correction, refund, or return. Anyone who received the product through a distributor rather than directly from Boston Scientific Corporation should confirm with their supplier whether their specific lot or unit is on the affected list, since recall notices are lot- or unit-specific and not every unit of a product line is necessarily included. Questions about a specific unit's status should go to Boston Scientific Corporation directly or to the distributor that supplied it; Argus HQ is a public-record aggregator, not the recalling firm, and cannot confirm individual unit status.
Boston Scientific Corporation’s FDA history
Argus HQ has recorded 294 total FDA actions tied to Boston Scientific Corporation: 0 warning letters, 294 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- Has Boston Scientific Corporation had other FDA recalls or enforcement actions?
- Yes. Argus HQ has recorded 294 total FDA actions tied to Boston Scientific Corporation, including 294 recalls and 0 warning letters.
- What is the FDA recall number for this event?
- FDA recall number Z-3082-2026.
- What is the current status of this recall?
- Status as recorded by FDA: Ongoing.
- Where can I verify this recall directly with FDA?
- The verbatim FDA record is the source of truth for every recall Argus HQ tracks — always confirm against fda.gov before acting.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Boston Scientific Corporation- Recall: Boston Scientific Corporation — Class I recall: Imager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M0… →
- Recall: Boston Scientific Corporation — Class II recall: Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power C… →
- Recall: Boston Scientific Corporation — Class II recall: LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT C… →
- Recall: Boston Scientific Corporation — Class II recall: LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT C… →
- Recall: Boston Scientific Corporation — Class I recall: Boston Scientific ENROUTE Transcarotid Neuroprotection System Plus, Material Nu… →
- Recall: Boston Scientific Corporation — Class I recall: Boston Scientific ENROUTE Transcarotid Neuroprotection System, Material Number… →
- Recall: Boston Scientific Corporation — Class II recall: Suture Cinch Long 1pk. Endoscopic tissue approximation device. →
- Recall: Boston Scientific Corporation — Class II recall: OverStitch Suture Cinch, 1pk. Endoscopic tissue approximation device. →
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2026). FDA Recall Database: Class II Recall: LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor… by Boston Scientific Corporation. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/recall/boston-scientific-corporation-z-3082-2026-2026-09-09
"FDA Recall Database: Class II Recall: LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor… by Boston Scientific Corporation." Argus HQ Research, Digital Empire Holdings LLC, 2026, argushq.ai/recall/boston-scientific-corporation-z-3082-2026-2026-09-09.
Argus HQ Research. "FDA Recall Database: Class II Recall: LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor… by Boston Scientific Corporation." Digital Empire Holdings LLC. Accessed October 12, 2026. https://argushq.ai/recall/boston-scientific-corporation-z-3082-2026-2026-09-09.
@misc{argushq_argushq_ai_recall_boston_scientific_corporation_z_3082_2026_2026_09_09_2026,
title = {FDA Recall Database: Class II Recall: LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor… by Boston Scientific Corporation},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/recall/boston-scientific-corporation-z-3082-2026-2026-09-09},
note = {Accessed: October 12, 2026}
}Source: FDA.gov — Recalls ↗