Key facts
| Recalling firm | Boston Scientific Neuromodulation Corporation |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Technique Manual |
| Classification | Class II |
| Recall number | Z-1972-2023 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury. |
What this classification means
A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote. FDA still requires the recalling firm to notify affected accounts and submit status reports on the correction, though the monitoring intensity is lower than for a Class I recall. FDA's Center for Devices and Radiological Health (for devices) or Center for Drug Evaluation and Research (for drugs) reviews the classification and can require additional corrective action if the initial response is judged insufficient. Because this record is filed as Class II, professionals and consumers handling an affected lot should treat FDA's own classification -- not this page's summary of it -- as the authoritative risk signal. A Class I recall is FDA’s most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case -- it reflects FDA’s assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field.
What consumers should do
Facilities, clinicians, and consumers holding an affected unit or lot of Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Technique Manual should stop using it for the purpose described in the recall notice and hold remaining stock pending instructions from Boston Scientific Neuromodulation Corporation. FDA recall records typically direct the recalling firm to notify known accounts directly with a recall letter describing the specific corrective action -- replacement, field correction, refund, or return. Anyone who received the product through a distributor rather than directly from Boston Scientific Neuromodulation Corporation should confirm with their supplier whether their specific lot or unit is on the affected list, since recall notices are lot- or unit-specific and not every unit of a product line is necessarily included. Questions about a specific unit's status should go to Boston Scientific Neuromodulation Corporation directly or to the distributor that supplied it; Argus HQ is a public-record aggregator, not the recalling firm, and cannot confirm individual unit status.
Boston Scientific Neuromodulation Corporation’s FDA history
Argus HQ has recorded 10 total FDA actions tied to Boston Scientific Neuromodulation Corporation: 0 warning letters, 10 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- What is the recall class for Boston Scientific Neuromodulation Corporation's recall?
- FDA classified this recall as Class II. A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
- Why was Driver Instrument, part of the Superion Indirect Decompressi recalled?
- FDA's recall record states the reason as: Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.
- Who should I contact about this recall?
- Contact Boston Scientific Neuromodulation Corporation directly, or the distributor/pharmacy that supplied the product, to confirm whether your specific lot or unit is affected. Argus HQ aggregates the public FDA recall record and is not the recalling firm.
- Has Boston Scientific Neuromodulation Corporation had other FDA recalls or enforcement actions?
- Yes. Argus HQ has recorded 10 total FDA actions tied to Boston Scientific Neuromodulation Corporation, including 10 recalls and 0 warning letters.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Boston Scientific Neuromodulation Corporation- Recall: Boston Scientific Neuromodulation Corporation — Class II recall: Intracept Access Instruments, Description/REF: INTRACEPT ACCESS INSTRUMENTS- DR… →
- Recall: Boston Scientific Neuromodulation Corporation — Class II recall: Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number:… →
- Recall: Boston Scientific Neuromodulation Corporation — Class II recall: Superion IDS Kit Superion Indirect Decompression System 8MM, VF Implant 101-98… →
- Recall: Boston Scientific Neuromodulation Corporation — Class II recall: Burr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB… →
- Recall: Boston Scientific Neuromodulation Corporation — Class II recall: WAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232,… →
- Recall: Boston Scientific Neuromodulation Corporation — Class II recall: Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16… →
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2023). FDA Class II Recall: Driver Instrument, part of the Superion Indirect Decompression… by Boston Scientific Neuromodulation Corporation — Recall Details. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/recall/boston-scientific-neuromodulation-corporation-z-1972-2023-2023-06-21
"FDA Class II Recall: Driver Instrument, part of the Superion Indirect Decompression… by Boston Scientific Neuromodulation Corporation — Recall Details." Argus HQ Research, Digital Empire Holdings LLC, 2023, argushq.ai/recall/boston-scientific-neuromodulation-corporation-z-1972-2023-2023-06-21.
Argus HQ Research. "FDA Class II Recall: Driver Instrument, part of the Superion Indirect Decompression… by Boston Scientific Neuromodulation Corporation — Recall Details." Digital Empire Holdings LLC. Accessed July 24, 2026. https://argushq.ai/recall/boston-scientific-neuromodulation-corporation-z-1972-2023-2023-06-21.
@misc{argushq_argushq_ai_recall_boston_scientific_neuromodulation_corporation_z_1972_2023_2023_06_21_2023,
title = {FDA Class II Recall: Driver Instrument, part of the Superion Indirect Decompression… by Boston Scientific Neuromodulation Corporation — Recall Details},
author = {{Argus HQ Research}},
year = {2023},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/recall/boston-scientific-neuromodulation-corporation-z-1972-2023-2023-06-21},
note = {Accessed: July 24, 2026}
}Source: FDA.gov — Recalls ↗