Key facts
| Recalling firm | Edwards Lifesciences, LLC |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | TruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm); PXVK0996:(3cc)/VAMP 68in (174cm); PXVMP2X21: x2 (3cc)/VAMP; PXVMP2X22: x2 (3cc)/VAMP 63 in (160cm); PXVMP3X32: x3 (3cc)/VAMP 75in (190cm); PXVP0041: (3cc)/VAMP Plus 96in (244cm); PXVP0550: x3 (3cc)/VAMP Plus 91 in (231cm); PXVP1071: x3 (3cc)/VAMP Plus 87in (221cm); PXVP1087: x4(3cc)/VAMP Plus; PXVP2260: (3cc)/VAMP Plus 60in (150cm); PXVP2272: (3cc)/VAMP Plus 72in (180cm); PXVP2284: (3cc)/VAMP Plus 84in (210cm); PXVP2284AT3: (3cc)/VAMP Plus 84in (210cm); PXVP23X3: x3: (3cc)/VAMP Plus; PXVP23X3AT3: x3 (3cc)/VAMP Plus; PXVP2X2AT3: x2 (3cc)/VAMP Plus; PXVPL0111: (3cc)/VAMP Plus 84in (213cm); PXVMP160: 3cc/VAMP 60 in (150cm); PXVMP172: 3cc/VAMP 72in)180cm); PXVMP184:3cc/VAMP 84in (210cm); PXVMP260:3cc/VAMP 60in (150cm); PXVFL10017: (3cc)/VAMP Flex 83in (210cm); PXVP0034: (3cc)/VAMP Plus 87in (220cm); T001691A: (3cc)/VAMP Plus; T001691M: (3cc)/VAMP Plus (215cm); T001744M:(3cc)/VAMP 93in (236 cm); T100671M: x3(3cc)/VAMP Plus (215cm); T001760A :x2(3cc)/VAMP Plus; T001765A: (3cc)/60in (150cm); T005021M: (3cc)/VAMP Flex (195cm); T005089M: x2 (3cc)/VAMP Flex (195 cm) ; T005091M: x3(3cc)/VAMP Flex (195cm); T434501B: x2(3cc)/VAMP 88in (225cm); T440B26B: x2 (3cc)/VAMP 88in (225cm); T450545A: x3 (3cc)/VAMP 75in (190cm); T460241A: (3cc)/VAMP Plus 69in (175cm); T530219D: x2 (3cc)/VAMP Plus 65in (165cm) |
| Classification | Class II |
| Recall number | Z-0414-2024 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer. |
What consumers should do
Facilities, clinicians, and consumers holding an affected unit or lot of TruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm); PXVK0996:(3cc)/VAMP 68in (174cm); PXVMP2X21: x2 (3cc)/VAMP; PXVMP2X22: x2 (3cc)/VAMP 63 in (160cm); PXVMP3X32: x3 (3cc)/VAMP 75in (190cm); PXVP0041: (3cc)/VAMP Plus 96in (244cm); PXVP0550: x3 (3cc)/VAMP Plus 91 in (231cm); PXVP1071: x3 (3cc)/VAMP Plus 87in (221cm); PXVP1087: x4(3cc)/VAMP Plus; PXVP2260: (3cc)/VAMP Plus 60in (150cm); PXVP2272: (3cc)/VAMP Plus 72in (180cm); PXVP2284: (3cc)/VAMP Plus 84in (210cm); PXVP2284AT3: (3cc)/VAMP Plus 84in (210cm); PXVP23X3: x3: (3cc)/VAMP Plus; PXVP23X3AT3: x3 (3cc)/VAMP Plus; PXVP2X2AT3: x2 (3cc)/VAMP Plus; PXVPL0111: (3cc)/VAMP Plus 84in (213cm); PXVMP160: 3cc/VAMP 60 in (150cm); PXVMP172: 3cc/VAMP 72in)180cm); PXVMP184:3cc/VAMP 84in (210cm); PXVMP260:3cc/VAMP 60in (150cm); PXVFL10017: (3cc)/VAMP Flex 83in (210cm); PXVP0034: (3cc)/VAMP Plus 87in (220cm); T001691A: (3cc)/VAMP Plus; T001691M: (3cc)/VAMP Plus (215cm); T001744M:(3cc)/VAMP 93in (236 cm); T100671M: x3(3cc)/VAMP Plus (215cm); T001760A :x2(3cc)/VAMP Plus; T001765A: (3cc)/60in (150cm); T005021M: (3cc)/VAMP Flex (195cm); T005089M: x2 (3cc)/VAMP Flex (195 cm) ; T005091M: x3(3cc)/VAMP Flex (195cm); T434501B: x2(3cc)/VAMP 88in (225cm); T440B26B: x2 (3cc)/VAMP 88in (225cm); T450545A: x3 (3cc)/VAMP 75in (190cm); T460241A: (3cc)/VAMP Plus 69in (175cm); T530219D: x2 (3cc)/VAMP Plus 65in (165cm) should stop using it for the purpose described in the recall notice and hold remaining stock pending instructions from Edwards Lifesciences, LLC. FDA recall records typically direct the recalling firm to notify known accounts directly with a recall letter describing the specific corrective action -- replacement, field correction, refund, or return. Anyone who received the product through a distributor rather than directly from Edwards Lifesciences, LLC should confirm with their supplier whether their specific lot or unit is on the affected list, since recall notices are lot- or unit-specific and not every unit of a product line is necessarily included. Questions about a specific unit's status should go to Edwards Lifesciences, LLC directly or to the distributor that supplied it; Argus HQ is a public-record aggregator, not the recalling firm, and cannot confirm individual unit status.
What this classification means
A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote. FDA still requires the recalling firm to notify affected accounts and submit status reports on the correction, though the monitoring intensity is lower than for a Class I recall. FDA's Center for Devices and Radiological Health (for devices) or Center for Drug Evaluation and Research (for drugs) reviews the classification and can require additional corrective action if the initial response is judged insufficient. Because this record is filed as Class II, professionals and consumers handling an affected lot should treat FDA's own classification -- not this page's summary of it -- as the authoritative risk signal. A Class I recall is FDA’s most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case -- it reflects FDA’s assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field.
Edwards Lifesciences, LLC’s FDA history
Argus HQ has recorded 17 total FDA actions tied to Edwards Lifesciences, LLC: 0 warning letters, 17 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- What is the current status of this recall?
- Status as recorded by FDA: Ongoing.
- Where can I verify this recall directly with FDA?
- The verbatim FDA record is the source of truth for every recall Argus HQ tracks — always confirm against fda.gov before acting.
- Is TruWave with Vamp System Pressure monitoring set, REF: PXA recalled?
- Yes. FDA lists a Class II recall by Edwards Lifesciences, LLC for TruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm); PXVK0996:(3cc)/VAMP 68in (174cm); PXVMP2X21: x2 (3cc)/VAMP; PXVMP2X22: x2 (3cc)/VAMP 63 in (160cm); PXVMP3X32: x3 (3cc)/VAMP 75in (190cm); PXVP0041: (3cc)/VAMP Plus 96in (244cm); PXVP0550: x3 (3cc)/VAMP Plus 91 in (231cm); PXVP1071: x3 (3cc)/VAMP Plus 87in (221cm); PXVP1087: x4(3cc)/VAMP Plus; PXVP2260: (3cc)/VAMP Plus 60in (150cm); PXVP2272: (3cc)/VAMP Plus 72in (180cm); PXVP2284: (3cc)/VAMP Plus 84in (210cm); PXVP2284AT3: (3cc)/VAMP Plus 84in (210cm); PXVP23X3: x3: (3cc)/VAMP Plus; PXVP23X3AT3: x3 (3cc)/VAMP Plus; PXVP2X2AT3: x2 (3cc)/VAMP Plus; PXVPL0111: (3cc)/VAMP Plus 84in (213cm); PXVMP160: 3cc/VAMP 60 in (150cm); PXVMP172: 3cc/VAMP 72in)180cm); PXVMP184:3cc/VAMP 84in (210cm); PXVMP260:3cc/VAMP 60in (150cm); PXVFL10017: (3cc)/VAMP Flex 83in (210cm); PXVP0034: (3cc)/VAMP Plus 87in (220cm); T001691A: (3cc)/VAMP Plus; T001691M: (3cc)/VAMP Plus (215cm); T001744M:(3cc)/VAMP 93in (236 cm); T100671M: x3(3cc)/VAMP Plus (215cm); T001760A :x2(3cc)/VAMP Plus; T001765A: (3cc)/60in (150cm); T005021M: (3cc)/VAMP Flex (195cm); T005089M: x2 (3cc)/VAMP Flex (195 cm) ; T005091M: x3(3cc)/VAMP Flex (195cm); T434501B: x2(3cc)/VAMP 88in (225cm); T440B26B: x2 (3cc)/VAMP 88in (225cm); T450545A: x3 (3cc)/VAMP 75in (190cm); T460241A: (3cc)/VAMP Plus 69in (175cm); T530219D: x2 (3cc)/VAMP Plus 65in (165cm), reported 20231206, with status "Ongoing." FDA recall number: Z-0414-2024.
- What is the recall class for Edwards Lifesciences, LLC's recall?
- FDA classified this recall as Class II. A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Edwards Lifesciences, LLC- Recall: Edwards Lifesciences, LLC — Class I recall: Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog N… →
- Recall: Edwards Lifesciences, LLC — Class I recall: Brand Name: FEM-FLEXII Product Name: Femoral Arterial Cannula Model/Catalog Nu… →
- Recall: Edwards Lifesciences, LLC — Class I recall: Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog N… →
- Recall: Edwards Lifesciences, LLC — Class I recall: Brand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog N… →
- Recall: Edwards Lifesciences, LLC — Class I recall: Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula… →
- Recall: Edwards Lifesciences, LLC — Class I recall: Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula… →
- Recall: Edwards Lifesciences, LLC — Class II recall: Acumen IQ Sensor with VAMP System, REFs: AIQS6AZ & AIQS6AZ5, STERILEEO, Rx only.… →
- Recall: Edwards Lifesciences, LLC — Class II recall: FloTrac Sensor, REF: MHD8 & MHD85,MHD6R5, MHD8R, MHD8R5, MHD8S, STERILEEO, Rx on… →
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2023). FDA Recall Database: Class II Recall: TruWave with Vamp System Pressure monitoring set, REF:… by Edwards Lifesciences, LLC. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/recall/edwards-lifesciences-z-0414-2024-2023-12-06
"FDA Recall Database: Class II Recall: TruWave with Vamp System Pressure monitoring set, REF:… by Edwards Lifesciences, LLC." Argus HQ Research, Digital Empire Holdings LLC, 2023, argushq.ai/recall/edwards-lifesciences-z-0414-2024-2023-12-06.
Argus HQ Research. "FDA Recall Database: Class II Recall: TruWave with Vamp System Pressure monitoring set, REF:… by Edwards Lifesciences, LLC." Digital Empire Holdings LLC. Accessed August 27, 2026. https://argushq.ai/recall/edwards-lifesciences-z-0414-2024-2023-12-06.
@misc{argushq_argushq_ai_recall_edwards_lifesciences_z_0414_2024_2023_12_06_2023,
title = {FDA Recall Database: Class II Recall: TruWave with Vamp System Pressure monitoring set, REF:… by Edwards Lifesciences, LLC},
author = {{Argus HQ Research}},
year = {2023},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/recall/edwards-lifesciences-z-0414-2024-2023-12-06},
note = {Accessed: August 27, 2026}
}Source: FDA.gov — Recalls ↗