Key facts
| Recalling firm | Emano Metrics, Inc. |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application. |
| Classification | Class II |
| Recall number | Z-2777-2026 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams. |
What this classification means
A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote. FDA still requires the recalling firm to notify affected accounts and submit status reports on the correction, though the monitoring intensity is lower than for a Class I recall. FDA's Center for Devices and Radiological Health (for devices) or Center for Drug Evaluation and Research (for drugs) reviews the classification and can require additional corrective action if the initial response is judged insufficient. Because this record is filed as Class II, professionals and consumers handling an affected lot should treat FDA's own classification -- not this page's summary of it -- as the authoritative risk signal. A Class I recall is FDA’s most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case -- it reflects FDA’s assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field.
What consumers should do
Facilities, clinicians, and consumers holding an affected unit or lot of Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application. should stop using it for the purpose described in the recall notice and hold remaining stock pending instructions from Emano Metrics, Inc.. FDA recall records typically direct the recalling firm to notify known accounts directly with a recall letter describing the specific corrective action -- replacement, field correction, refund, or return. Anyone who received the product through a distributor rather than directly from Emano Metrics, Inc. should confirm with their supplier whether their specific lot or unit is on the affected list, since recall notices are lot- or unit-specific and not every unit of a product line is necessarily included. Questions about a specific unit's status should go to Emano Metrics, Inc. directly or to the distributor that supplied it; Argus HQ is a public-record aggregator, not the recalling firm, and cannot confirm individual unit status.
Emano Metrics, Inc.’s FDA history
Argus HQ has recorded 1 total FDA action tied to Emano Metrics, Inc.: 0 warning letters, 1 recall, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- Is Emano Flow with software version v2.0.0 GS1 GTIN code: 1998 recalled?
- Yes. FDA lists a Class II recall by Emano Metrics, Inc. for Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application., reported 20260729, with status "Ongoing." FDA recall number: Z-2777-2026.
- What is the recall class for Emano Metrics, Inc.'s recall?
- FDA classified this recall as Class II. A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
- Why was Emano Flow with software version v2.0.0 GS1 GTIN code: 1998 recalled?
- FDA's recall record states the reason as: Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.
- Who should I contact about this recall?
- Contact Emano Metrics, Inc. directly, or the distributor/pharmacy that supplied the product, to confirm whether your specific lot or unit is affected. Argus HQ aggregates the public FDA recall record and is not the recalling firm.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Emano Metrics, Inc.Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2026). FDA Recall Database: Class II Recall: Emano Flow with software version v2.0.0 GS1 GTIN… by Emano Metrics, Inc.. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/recall/emano-metrics-z-2777-2026-2026-07-29
"FDA Recall Database: Class II Recall: Emano Flow with software version v2.0.0 GS1 GTIN… by Emano Metrics, Inc.." Argus HQ Research, Digital Empire Holdings LLC, 2026, argushq.ai/recall/emano-metrics-z-2777-2026-2026-07-29.
Argus HQ Research. "FDA Recall Database: Class II Recall: Emano Flow with software version v2.0.0 GS1 GTIN… by Emano Metrics, Inc.." Digital Empire Holdings LLC. Accessed August 4, 2026. https://argushq.ai/recall/emano-metrics-z-2777-2026-2026-07-29.
@misc{argushq_argushq_ai_recall_emano_metrics_z_2777_2026_2026_07_29_2026,
title = {FDA Recall Database: Class II Recall: Emano Flow with software version v2.0.0 GS1 GTIN… by Emano Metrics, Inc.},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/recall/emano-metrics-z-2777-2026-2026-07-29},
note = {Accessed: August 4, 2026}
}Source: FDA.gov — Recalls ↗