Key facts
| Recalling firm | ARROW INTERNATIONAL, LLC |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | Description/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A; AM/AT PI MIDLINE 1L: 4.5FR X 15CM/CDC-41541-MPK1A; AM/AT PI MIDLINE 2L: 5.5FR X 15CM/CDC-41552-MPK1A; AGBA PI MIDLINE 1L: 4.5FR X 15CM/ASK-41541-VFHM; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPK; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPKB; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPKC; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPK; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPKB; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPKC; PI MIDLINE 1-L: 4FR X 20CM/CDC-32041-MPK; PI MIDLINE 1-L: 4FR X 20CM/CDC-32041-MPKB; PI MIDLINE 1L: 4FR X 20CM/CDC-32041-MPKC; PI MIDLINE 2-L: 5FR X 20CM/CDC-32052-MPKB; PI PICC NaviCurve: 1L 4FR x 55CM w BP/DLX-35541-CURVB; PI PICC: 1L 4FR x 55CM w Biopatch/DLX-35541-HPKB; PI PICC: 1L 4FR x 55CM w CHG/DLX-35541-HPKC; PI PICC: 2L 5FR x 55CM w Biopatch/DLX-35552-HPKB; PI PICC: 2L 5FR X 55CM w CHG/DLX-35552-HPKC; PI PICC: 3L 6FR x 55CM w Biopatch/DLX-35563-HPKB; AGBA PICC/DELTA KIT: 1L 4.5 FR X 40 CM/ASK-44041-NS; AGBA PICC/DELTA FG: 2L 5.5 FR X 40 CM/ASK-44052-NS; AGBA PICC/DELTA KIT: 1L 4.5 FR X 50 CM/ASK-45041-NS; PI PICC KIT: 2L 5.5 FR X 50 CM/ASK-45052-TG; PI AGBA PICC KIT: 3L 6 FR X 50 CM W/VPS/ASK-45063-NS; PICC KIT: 1L 4.5 FR X 55 CM CG+ VPS/ASK-45541-RH; AGBA PICC: 1L 4.5FR x 55CM w Teg CHG/DLX-45541-HPKC; AGBA PICC: 2L 5.5FR x 55CM w Teg CHG/DLX-45552-HPKC; AGBA PICC NaviCurve: 3L 6FR x 55CM TCG/DLX-45563-CURVC; AGBA PICC: 3L 6FR x 55CM w Biopatch/DLX-45563-HPKB; AGBA PICC: 3L 6FR x 55CM w Teg CHG/DLX-45563-HPKC; PI PICC KIT: 1-L 4 FR X 55 CM TIPTRACKER/CDC-35541-TTS; PI PICC NaviCurve: 1L 4FR x 55CM w CHG/DLX-35541-CURVC; PI PICC G4 Stylet: 1L 4FR x 55CM w BP/DLX-35541-VPSB; PI PICC NaviCurve: 2L 5FR x 55CM w CHG/DLX-35552-CURVC; PI PICC G4 Stylet: 2L 5FR x 55CM w BP/DLX-35552-VPSB; PI PICC G4 Stylet: 2L 5FR x 55CM w CHG/DLX-35552-VPSC; PI PICC NaviCurve: 3L 6FR x 55CM w CHG/DLX-35563-CURVC; PI PICC G4 Stylet: 3L 6FR x 55CM w CHG/DLX-35563-VPSC; AGBA PICC/DELTA KIT: 1-L 4.5 FR X 40 CM/CDC-44041-VPS2; PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-UWH1; PI AGBA PICC KIT: 2L 5.5FR X 55CM W/ VPS/ASK-45552-UWH1; AGBA PICC G4 Stylet: 1L 4.5FR x 55CM BP/DLX-45541-VPSB; ACCESS TRAY/ASK-04001-NWM; ACCESS TRAY/ASK-04001-VCUH3; PI PICC:1L 4FR X 40CM PRELOAD VPS STYLET/CDC-34041-VPS; PI PICC:2L 5FR X 50CM PRELOAD VPS STYLET/CDC-35052-VPS; ACCESS KIT/ASK-04001-BWK2; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-DU11; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-PVD; MULTI LUMEN PI CVC KIT: 3L 7 FR X 16CM/ASK-12703-LMDTG1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8IN/ASK-15703-ECMC4; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-15703-PNM; CVC KIT: 3 LUMEN 12 FR X 6IN/ASK-22123-UAB; 2-L CVC KIT: 4 FR X13 CM/CDC-24402-X1A; LBCVC KIT: 3L 12 FR X 20 CM AGB/CDC-25123-X1A; CVC KIT: 3L 7 FR X 16 CM/ASK-42703-CPR; CVC KIT: 2L 8 FR X 6 IN (16 CM) AGB+/ASK-42802-DMC2; CVC KIT: 2 LUMEN 8 FR X 8 IN (20 CM)/ASK-45802-PCMH1; PI CVC KIT: 3L 7 FR X 6 IN (16 CM) AGB+/ASK-42703-MM1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PHUMC1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PMSG1; MULTILUMEN PI CVC KIT: 3L 7 FR X 6/ASK-42703-POSU3; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PSAU1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-RHR1; PI CATHETER 2L: 8 FR X 16 CM AGB+/ASK-42802-DU1; PI CVC KIT: 2L 8 FR X 6 IN/ASK-42802-PHF4; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-PHHN1; MULTILUMEN CVC KIT: 2L 8 FR X 6 IN/ASK-42802-POSU3; PI CVC KIT: 2-L 8 FR X 16 CM AGB+/ASK-42802-PUCD1; MULTI-LUMEN CVC KIT:2-L 8FR X 6IN (16CM)/ASK-42802-PWHC1; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-UHC; PI CVC KIT: 4L 8.5 FR X 16 CM AGB+/ASK-42854-MCY1; PI CVC KIT: 3L 7FR X 8IN (20 CM) AGB+/ASK-45703-BCH; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-EMC; PI CVC KIT: 3 LUMEN 7 FR X 8 IN (20 CM)/ASK-45703-MHMC1; PI CVC KIT: 3-L 7 FR X 20 CM AGB+/ASK-45703-NWM; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PCOR; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PIMH; MULTI LUMEN CVC KIT: 3L |
| Classification | Class I |
| Recall number | Z-2394-2026 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries. |
What consumers should do
Facilities and clinicians holding an affected lot or unit should stop using it for the indication described in the recall notice and quarantine remaining stock pending instructions from ARROW INTERNATIONAL, LLC. FDA recall records for devices typically direct the recalling firm to notify known accounts (hospitals, distributors, clinics) directly with a recall letter describing the specific corrective action — replacement, field correction, or return. Anyone who received Description/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A; AM/AT PI MIDLINE 1L: 4.5FR X 15CM/CDC-41541-MPK1A; AM/AT PI MIDLINE 2L: 5.5FR X 15CM/CDC-41552-MPK1A; AGBA PI MIDLINE 1L: 4.5FR X 15CM/ASK-41541-VFHM; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPK; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPKB; AGBA PI MIDLINE 1-L: 4.5FR X 15CM/CDC-41541-MPKC; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPK; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPKB; AGBA PI MIDLINE 2-L: 5.5FR X 15CM/CDC-41552-MPKC; PI MIDLINE 1-L: 4FR X 20CM/CDC-32041-MPK; PI MIDLINE 1-L: 4FR X 20CM/CDC-32041-MPKB; PI MIDLINE 1L: 4FR X 20CM/CDC-32041-MPKC; PI MIDLINE 2-L: 5FR X 20CM/CDC-32052-MPKB; PI PICC NaviCurve: 1L 4FR x 55CM w BP/DLX-35541-CURVB; PI PICC: 1L 4FR x 55CM w Biopatch/DLX-35541-HPKB; PI PICC: 1L 4FR x 55CM w CHG/DLX-35541-HPKC; PI PICC: 2L 5FR x 55CM w Biopatch/DLX-35552-HPKB; PI PICC: 2L 5FR X 55CM w CHG/DLX-35552-HPKC; PI PICC: 3L 6FR x 55CM w Biopatch/DLX-35563-HPKB; AGBA PICC/DELTA KIT: 1L 4.5 FR X 40 CM/ASK-44041-NS; AGBA PICC/DELTA FG: 2L 5.5 FR X 40 CM/ASK-44052-NS; AGBA PICC/DELTA KIT: 1L 4.5 FR X 50 CM/ASK-45041-NS; PI PICC KIT: 2L 5.5 FR X 50 CM/ASK-45052-TG; PI AGBA PICC KIT: 3L 6 FR X 50 CM W/VPS/ASK-45063-NS; PICC KIT: 1L 4.5 FR X 55 CM CG+ VPS/ASK-45541-RH; AGBA PICC: 1L 4.5FR x 55CM w Teg CHG/DLX-45541-HPKC; AGBA PICC: 2L 5.5FR x 55CM w Teg CHG/DLX-45552-HPKC; AGBA PICC NaviCurve: 3L 6FR x 55CM TCG/DLX-45563-CURVC; AGBA PICC: 3L 6FR x 55CM w Biopatch/DLX-45563-HPKB; AGBA PICC: 3L 6FR x 55CM w Teg CHG/DLX-45563-HPKC; PI PICC KIT: 1-L 4 FR X 55 CM TIPTRACKER/CDC-35541-TTS; PI PICC NaviCurve: 1L 4FR x 55CM w CHG/DLX-35541-CURVC; PI PICC G4 Stylet: 1L 4FR x 55CM w BP/DLX-35541-VPSB; PI PICC NaviCurve: 2L 5FR x 55CM w CHG/DLX-35552-CURVC; PI PICC G4 Stylet: 2L 5FR x 55CM w BP/DLX-35552-VPSB; PI PICC G4 Stylet: 2L 5FR x 55CM w CHG/DLX-35552-VPSC; PI PICC NaviCurve: 3L 6FR x 55CM w CHG/DLX-35563-CURVC; PI PICC G4 Stylet: 3L 6FR x 55CM w CHG/DLX-35563-VPSC; AGBA PICC/DELTA KIT: 1-L 4.5 FR X 40 CM/CDC-44041-VPS2; PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-UWH1; PI AGBA PICC KIT: 2L 5.5FR X 55CM W/ VPS/ASK-45552-UWH1; AGBA PICC G4 Stylet: 1L 4.5FR x 55CM BP/DLX-45541-VPSB; ACCESS TRAY/ASK-04001-NWM; ACCESS TRAY/ASK-04001-VCUH3; PI PICC:1L 4FR X 40CM PRELOAD VPS STYLET/CDC-34041-VPS; PI PICC:2L 5FR X 50CM PRELOAD VPS STYLET/CDC-35052-VPS; ACCESS KIT/ASK-04001-BWK2; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-DU11; MULTI LUMEN PI CVC KIT: 3L 7 FR X 6 IN/ASK-42703-PVD; MULTI LUMEN PI CVC KIT: 3L 7 FR X 16CM/ASK-12703-LMDTG1; MULTI LUMEN PI CVC KIT: 3L 7 FR X 8IN/ASK-15703-ECMC4; MULTI LUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-15703-PNM; CVC KIT: 3 LUMEN 12 FR X 6IN/ASK-22123-UAB; 2-L CVC KIT: 4 FR X13 CM/CDC-24402-X1A; LBCVC KIT: 3L 12 FR X 20 CM AGB/CDC-25123-X1A; CVC KIT: 3L 7 FR X 16 CM/ASK-42703-CPR; CVC KIT: 2L 8 FR X 6 IN (16 CM) AGB+/ASK-42802-DMC2; CVC KIT: 2 LUMEN 8 FR X 8 IN (20 CM)/ASK-45802-PCMH1; PI CVC KIT: 3L 7 FR X 6 IN (16 CM) AGB+/ASK-42703-MM1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PHUMC1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PMSG1; MULTILUMEN PI CVC KIT: 3L 7 FR X 6/ASK-42703-POSU3; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PSAU1; PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-RHR1; PI CATHETER 2L: 8 FR X 16 CM AGB+/ASK-42802-DU1; PI CVC KIT: 2L 8 FR X 6 IN/ASK-42802-PHF4; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-PHHN1; MULTILUMEN CVC KIT: 2L 8 FR X 6 IN/ASK-42802-POSU3; PI CVC KIT: 2-L 8 FR X 16 CM AGB+/ASK-42802-PUCD1; MULTI-LUMEN CVC KIT:2-L 8FR X 6IN (16CM)/ASK-42802-PWHC1; PI CVC KIT: 2L 8 FR X 16 CM AGB+/ASK-42802-UHC; PI CVC KIT: 4L 8.5 FR X 16 CM AGB+/ASK-42854-MCY1; PI CVC KIT: 3L 7FR X 8IN (20 CM) AGB+/ASK-45703-BCH; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-EMC; PI CVC KIT: 3 LUMEN 7 FR X 8 IN (20 CM)/ASK-45703-MHMC1; PI CVC KIT: 3-L 7 FR X 20 CM AGB+/ASK-45703-NWM; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PCOR; MULTILUMEN PI CVC KIT: 3L 7 FR X 20 CM/ASK-45703-PIMH; MULTI LUMEN CVC KIT: 3L through a distributor rather than directly from ARROW INTERNATIONAL, LLC should still confirm with their supplier whether their specific lot or serial number is on the affected list, since recall notices are lot- or unit-specific and not every unit of a product line is necessarily included. Questions about a specific unit's status should go to ARROW INTERNATIONAL, LLC directly or to the distributor that supplied it; Argus HQ is a public-record aggregator, not the recalling firm, and cannot confirm individual unit status.
What this classification means
A Class I recall is FDA's most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case — it reflects FDA's assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field. FDA's Center for Devices and Radiological Health (for devices) or Center for Drug Evaluation and Research (for drugs) reviews the classification and can require additional corrective action if the initial response is judged insufficient. Because Class I is reserved for the highest-risk defects, professionals and consumers handling an affected lot are generally advised to treat the recall as time-sensitive rather than routine.
ARROW INTERNATIONAL, LLC’s FDA history
Argus HQ has recorded 47 total FDA actions tied to ARROW INTERNATIONAL, LLC: 0 warning letters, 47 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- Why was Description/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-0203 recalled?
- FDA's recall record states the reason as: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if ster
- Who should I contact about this recall?
- Contact ARROW INTERNATIONAL, LLC directly, or the distributor/pharmacy that supplied the product, to confirm whether your specific lot or unit is affected. Argus HQ aggregates the public FDA recall record and is not the recalling firm.
- Has ARROW INTERNATIONAL, LLC had other FDA recalls or enforcement actions?
- Yes. Argus HQ has recorded 47 total FDA actions tied to ARROW INTERNATIONAL, LLC, including 47 recalls and 0 warning letters.
- What is the FDA recall number for this event?
- FDA recall number Z-2394-2026.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for ARROW INTERNATIONAL, LLCCompiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2026). Recall Record: Class I Recall: Description/REF: MIDLINE CATHETER KIT: 3 FR X 20… by ARROW INTERNATIONAL, LLC. Digital Empire LLC. Retrieved from https://argushq.ai/recall/fda-recall-arrow-international-llc-z-2394-2026
"Recall Record: Class I Recall: Description/REF: MIDLINE CATHETER KIT: 3 FR X 20… by ARROW INTERNATIONAL, LLC." Argus HQ Research, Digital Empire LLC, 2026, argushq.ai/recall/fda-recall-arrow-international-llc-z-2394-2026.
Argus HQ Research. "Recall Record: Class I Recall: Description/REF: MIDLINE CATHETER KIT: 3 FR X 20… by ARROW INTERNATIONAL, LLC." Digital Empire LLC. Accessed July 13, 2026. https://argushq.ai/recall/fda-recall-arrow-international-llc-z-2394-2026.
@misc{argushq_argushq_ai_recall_fda_recall_arrow_international_llc_z_2394_2026_2026,
title = {Recall Record: Class I Recall: Description/REF: MIDLINE CATHETER KIT: 3 FR X 20… by ARROW INTERNATIONAL, LLC},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/recall/fda-recall-arrow-international-llc-z-2394-2026},
note = {Accessed: July 13, 2026}
}Source: FDA.gov — Recalls ↗