Key facts
| Recalling firm | Cook Incorporated |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF). |
| Classification | Class I |
| Recall number | Z-1935-2026 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death. |
What consumers should do
Because FDA has assigned this recall its highest severity classification, anyone who has this product should stop using it immediately, regardless of whether a problem has been noticed yet. Specifically, check whether the product in your possession matches the description on file with FDA -- Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip (additional items listed in FDA's full record) -- and the recall number Z-1935-2026. If it matches, stop using or distributing it and contact Cook Incorporated or the point of purchase for return or replacement instructions. Retailers and distributors holding affected inventory should quarantine it and follow the firm's written recall notice.
What this classification means
FDA classifies a recall as Class I when there is a reasonable probability that using the recalled product, or being exposed to it, will cause serious adverse health consequences or death. Class I is FDA's most serious recall classification. In this case, the recalling firm is Cook Incorporated and the affected product is described in FDA's record as: Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip (additional items listed in FDA's full record). FDA's stated reason for the recall is: Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.. The recall is tracked under FDA recall number Z-1935-2026 and carries a status of Ongoing as of the most recent FDA enforcement report. This classification reflects FDA's assessment of the product defect described above, not a prediction about any specific patient outcome. FDA maintains a public enforcement report database so that consumers, physicians, and purchasers can look up the exact product identifiers involved before deciding what to do.
Cook Incorporated’s FDA history
Argus HQ has recorded 21 total FDA actions tied to Cook Incorporated: 0 warning letters, 21 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- What should someone who has this product do?
- Because FDA has assigned this recall its highest severity classification, anyone who has this product should stop using it immediately, regardless of whether a problem has been noticed yet. Check the product against FDA's description (Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip (additional items listed in FDA's full record)) and recall number.
- Has Cook Incorporated had other FDA recalls or enforcement actions?
- Yes. Argus HQ has recorded 21 total FDA actions tied to Cook Incorporated, including 21 recalls and 0 warning letters.
- What is the FDA recall number for this event?
- FDA recall number Z-1935-2026.
- What is the current status of this recall?
- Status as recorded by FDA: Ongoing.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Cook IncorporatedCompiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2026). FDA Class I Recall: Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308… by Cook Incorporated. Digital Empire LLC. Retrieved from https://argushq.ai/recall/fda-recall-cook-incorporated-z-1935-2026
"FDA Class I Recall: Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308… by Cook Incorporated." Argus HQ Research, Digital Empire LLC, 2026, argushq.ai/recall/fda-recall-cook-incorporated-z-1935-2026.
Argus HQ Research. "FDA Class I Recall: Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308… by Cook Incorporated." Digital Empire LLC. Accessed July 13, 2026. https://argushq.ai/recall/fda-recall-cook-incorporated-z-1935-2026.
@misc{argushq_argushq_ai_recall_fda_recall_cook_incorporated_z_1935_2026_2026,
title = {FDA Class I Recall: Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308… by Cook Incorporated},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/recall/fda-recall-cook-incorporated-z-1935-2026},
note = {Accessed: July 13, 2026}
}Source: FDA.gov — Recalls ↗