Key facts
| Recalling firm | Datex-Ohmeda Inc |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance. |
| Classification | Class I |
| Recall number | Z-2358-2026 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient. |
What this classification means
A Class I recall is FDA's most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case — it reflects FDA's assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field. FDA's Center for Devices and Radiological Health (for devices) or Center for Drug Evaluation and Research (for drugs) reviews the classification and can require additional corrective action if the initial response is judged insufficient. Because Class I is reserved for the highest-risk defects, professionals and consumers handling an affected lot are generally advised to treat the recall as time-sensitive rather than routine.
What consumers should do
Facilities and clinicians holding an affected lot or unit should stop using it for the indication described in the recall notice and quarantine remaining stock pending instructions from Datex-Ohmeda Inc. FDA recall records for devices typically direct the recalling firm to notify known accounts (hospitals, distributors, clinics) directly with a recall letter describing the specific corrective action — replacement, field correction, or return. Anyone who received Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance. through a distributor rather than directly from Datex-Ohmeda Inc should still confirm with their supplier whether their specific lot or serial number is on the affected list, since recall notices are lot- or unit-specific and not every unit of a product line is necessarily included. Questions about a specific unit's status should go to Datex-Ohmeda Inc directly or to the distributor that supplied it; Argus HQ is a public-record aggregator, not the recalling firm, and cannot confirm individual unit status.
Datex-Ohmeda Inc’s FDA history
Argus HQ has recorded 3 total FDA actions tied to Datex-Ohmeda Inc: 0 warning letters, 3 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- Is Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with In recalled?
- Yes. FDA lists a Class I recall by Datex-Ohmeda Inc for Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance., reported June 24, 2026, with status "Ongoing." FDA recall number: Z-2358-2026.
- What is the recall class for Datex-Ohmeda Inc's Giraffe Warmer with Integrated Resuscitation System (iRes), recall?
- FDA classified this recall as Class I. Class I is FDA's most serious classification, reserved for defects with a reasonable probability of serious health consequences or death.
- Why was Giraffe Warmer with Integrated Resuscitation System (iRes), recalled?
- FDA's recall record states the reason as: Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in
- Who should I contact about this recall?
- Contact Datex-Ohmeda Inc directly, or the distributor/pharmacy that supplied the product, to confirm whether your specific lot or unit is affected. Argus HQ aggregates the public FDA recall record and is not the recalling firm.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Datex-Ohmeda IncCompiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2026). Recall Record: Class I Recall: Giraffe Warmer with Integrated Resuscitation System (iRes), Panda… by Datex-Ohmeda Inc. Digital Empire LLC. Retrieved from https://argushq.ai/recall/fda-recall-datex-ohmeda-inc-z-2358-2026
"Recall Record: Class I Recall: Giraffe Warmer with Integrated Resuscitation System (iRes), Panda… by Datex-Ohmeda Inc." Argus HQ Research, Digital Empire LLC, 2026, argushq.ai/recall/fda-recall-datex-ohmeda-inc-z-2358-2026.
Argus HQ Research. "Recall Record: Class I Recall: Giraffe Warmer with Integrated Resuscitation System (iRes), Panda… by Datex-Ohmeda Inc." Digital Empire LLC. Accessed July 13, 2026. https://argushq.ai/recall/fda-recall-datex-ohmeda-inc-z-2358-2026.
@misc{argushq_argushq_ai_recall_fda_recall_datex_ohmeda_inc_z_2358_2026_2026,
title = {Recall Record: Class I Recall: Giraffe Warmer with Integrated Resuscitation System (iRes), Panda… by Datex-Ohmeda Inc},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/recall/fda-recall-datex-ohmeda-inc-z-2358-2026},
note = {Accessed: July 13, 2026}
}Source: FDA.gov — Recalls ↗