Key facts
| Recalling firm | MICROVENTION INC. |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm |
| Classification | Class II |
| Recall number | Z-2200-2026 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition, acute stent thrombosis, distal embolization, ischemic neurological events |
What this classification means
FDA classifies a recall as Class II when using the recalled product, or being exposed to it, may cause temporary or medically reversible adverse health consequences, or when the probability of serious adverse health consequences is remote. In this case, the recalling firm is MICROVENTION INC. and the affected product is described in FDA's record as: LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm. FDA's stated reason for the recall is: Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition (additional items listed in FDA's full record). The recall is tracked under FDA recall number Z-2200-2026 and carries a status of Ongoing as of the most recent FDA enforcement report. This classification reflects FDA's assessment of the product defect described above, not a prediction about any specific patient outcome. FDA maintains a public enforcement report database so that consumers, physicians, and purchasers can look up the exact product identifiers involved before deciding what to do.
What consumers should do
Because this recall carries FDA's middle severity classification, anyone who has this product should stop using it and follow the recalling firm's instructions, even though serious harm is considered unlikely for this classification. Specifically, check whether the product in your possession matches the description on file with FDA -- LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm -- and the recall number Z-2200-2026. If it matches, stop using or distributing it and contact MICROVENTION INC. or the point of purchase for return or replacement instructions. Retailers and distributors holding affected inventory should quarantine it and follow the firm's written recall notice.
MICROVENTION INC.’s FDA history
Argus HQ has recorded 1 total FDA action tied to MICROVENTION INC.: 0 warning letters, 1 recall, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- Where can I verify this recall directly with FDA?
- The verbatim FDA record is the source of truth for every recall Argus HQ tracks — always confirm against fda.gov before acting.
- What was recalled and why?
- MICROVENTION INC. recalled LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm. FDA's record states the reason for recall as: Stent-assisted coiling device may due to commingling during manufacturing have dimensional nonconformance and/or incorrect device configuration. If device is oversized may lead to increased vessel wall stress, neointimal hyperplasia, vessel injury/rupture; if undersized may lead to incomplete expansion/malposition (additional items listed.
- How serious is this recall?
- FDA classifies a recall as Class II when using the recalled product, or being exposed to it, may cause temporary or medically reversible adverse health consequences, or when the probability of serious adverse health consequences is remote.
- What should someone who has this product do?
- Because this recall carries FDA's middle severity classification, anyone who has this product should stop using it and follow the recalling firm's instructions, even though serious harm is considered unlikely for this classification. Check the product against FDA's description (LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL): 214022-CAS/4.0 mm x 22 mm x 18 mm, 214518-CAS/4.5 mm x 18 mm x 14 mm, 213517-CAS/3.5 mm x 17 mm x 13 mm) and recall number Z-2200-2026, then follow MICROVENTION INC.'s recall instructions.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for MICROVENTION INC.Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2026). Recall Record: Class II Recall: LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL):… by MICROVENTION INC.. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/recall/fda-recall-microvention-inc-z-2200-2026
"Recall Record: Class II Recall: LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL):… by MICROVENTION INC.." Argus HQ Research, Digital Empire Holdings LLC, 2026, argushq.ai/recall/fda-recall-microvention-inc-z-2200-2026.
Argus HQ Research. "Recall Record: Class II Recall: LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL):… by MICROVENTION INC.." Digital Empire Holdings LLC. Accessed August 13, 2026. https://argushq.ai/recall/fda-recall-microvention-inc-z-2200-2026.
@misc{argushq_argushq_ai_recall_fda_recall_microvention_inc_z_2200_2026_2026,
title = {Recall Record: Class II Recall: LVIS Intraluminal Support Device, REF/Outside Diameter(OD)xTotal Length(TL)xWorking Length(WL):… by MICROVENTION INC.},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/recall/fda-recall-microvention-inc-z-2200-2026},
note = {Accessed: August 13, 2026}
}Source: FDA.gov — Recalls ↗