Key facts
| Recalling firm | Medtronic Perfusion Systems |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | MC3 VitalFlow Console, REF 58100; Blood pump of ecmo |
| Classification | Class II |
| Recall number | Z-0509-2026 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout. |
What consumers should do
Facilities, clinicians, and consumers holding an affected unit or lot of MC3 VitalFlow Console, REF 58100; Blood pump of ecmo should stop using it for the purpose described in the recall notice and hold remaining stock pending instructions from Medtronic Perfusion Systems. FDA recall records typically direct the recalling firm to notify known accounts directly with a recall letter describing the specific corrective action -- replacement, field correction, refund, or return. Anyone who received the product through a distributor rather than directly from Medtronic Perfusion Systems should confirm with their supplier whether their specific lot or unit is on the affected list, since recall notices are lot- or unit-specific and not every unit of a product line is necessarily included. Questions about a specific unit's status should go to Medtronic Perfusion Systems directly or to the distributor that supplied it; Argus HQ is a public-record aggregator, not the recalling firm, and cannot confirm individual unit status.
What this classification means
A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote. FDA still requires the recalling firm to notify affected accounts and submit status reports on the correction, though the monitoring intensity is lower than for a Class I recall. FDA's Center for Devices and Radiological Health (for devices) or Center for Drug Evaluation and Research (for drugs) reviews the classification and can require additional corrective action if the initial response is judged insufficient. Because this record is filed as Class II, professionals and consumers handling an affected lot should treat FDA's own classification -- not this page's summary of it -- as the authoritative risk signal. A Class I recall is FDA’s most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case -- it reflects FDA’s assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field.
Medtronic Perfusion Systems’s FDA history
Argus HQ has recorded 46 total FDA actions tied to Medtronic Perfusion Systems: 0 warning letters, 46 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- Why was MC3 VitalFlow Console, REF 58100; Blood pump of ecmo recalled?
- FDA's recall record states the reason as: As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.
- Who should I contact about this recall?
- Contact Medtronic Perfusion Systems directly, or the distributor/pharmacy that supplied the product, to confirm whether your specific lot or unit is affected. Argus HQ aggregates the public FDA recall record and is not the recalling firm.
- Has Medtronic Perfusion Systems had other FDA recalls or enforcement actions?
- Yes. Argus HQ has recorded 46 total FDA actions tied to Medtronic Perfusion Systems, including 46 recalls and 0 warning letters.
- What is the FDA recall number for this event?
- FDA recall number Z-0509-2026.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Medtronic Perfusion Systems- Recall: Medtronic Perfusion Systems — Class II recall: Octopus Nuvo Tissue Stabilizer, disposable tissue stabilization system, Product… →
- Recall: Medtronic Perfusion Systems — Class II recall: Medtronic HR-ACT (High Range Activated Clotting Time Cartridges), REF 402-03 →
- Recall: Medtronic Perfusion Systems — Class II recall: Affinity NT Oxygenator, Oxygenator with Balance Biosurface and Uncoated Cardioto… →
- Recall: Medtronic Perfusion Systems — Class I recall: DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12110… →
- Recall: Medtronic Perfusion Systems — Class I recall: DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12115… →
- Recall: Medtronic Perfusion Systems — Class I recall: DLP Left Heart Vent Catheter Malleable body and vented connector, SUCTION 12113… →
- Recall: Medtronic Perfusion Systems — Class I recall: Aortic Root Cannula: 1. DLP Aortic Root Cannula A. CANNULA 11012 AR FLOWGUARD… →
- Recall: Medtronic Perfusion Systems — Class II recall: Select Series Angled Tip Arterial Cannula, Model Number 72422 (22 Fr) →
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2025). Recall Record: Class II Recall: MC3 VitalFlow Console, REF 58100; Blood pump of… by Medtronic Perfusion Systems. Digital Empire Holdings LLC. Retrieved from https://argushq.ai/recall/medtronic-perfusion-systems-z-0509-2026-2025-11-26
"Recall Record: Class II Recall: MC3 VitalFlow Console, REF 58100; Blood pump of… by Medtronic Perfusion Systems." Argus HQ Research, Digital Empire Holdings LLC, 2025, argushq.ai/recall/medtronic-perfusion-systems-z-0509-2026-2025-11-26.
Argus HQ Research. "Recall Record: Class II Recall: MC3 VitalFlow Console, REF 58100; Blood pump of… by Medtronic Perfusion Systems." Digital Empire Holdings LLC. Accessed July 24, 2026. https://argushq.ai/recall/medtronic-perfusion-systems-z-0509-2026-2025-11-26.
@misc{argushq_argushq_ai_recall_medtronic_perfusion_systems_z_0509_2026_2025_11_26_2025,
title = {Recall Record: Class II Recall: MC3 VitalFlow Console, REF 58100; Blood pump of… by Medtronic Perfusion Systems},
author = {{Argus HQ Research}},
year = {2025},
publisher = {Digital Empire Holdings LLC},
url = {https://argushq.ai/recall/medtronic-perfusion-systems-z-0509-2026-2025-11-26},
note = {Accessed: July 24, 2026}
}Source: FDA.gov — Recalls ↗