Key facts
| Recalling firm | Roche Diagnostics Operations, Inc. |
|---|---|
| Manufacturer | Not disclosed in the FDA record. |
| Brand name | Not disclosed in the FDA record. |
| Product description | cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001 cobas pro SSU, material number 09205632001 Affected Assays: (Product Name/ Catalog Number/ Application Code/ Application Short Name) 1. Cystatin C Gen.2 08105596190 20510 CYSC2 2. Ferritin Gen.4 08057648190 20571 FER4X 3. Lipoprotein (a) Gen.2 08106126190 20861 LPA2X 4. Lipoprotein (a) molarity 08106126160 20864 LPA2-X 5. Vancomycin Gen.3 08058849190 21211 VANC3O 6. Kappa Free Light Chains Partner Channel 08896640190 21421 KFLCX 7. Lambda Free Light Chains Partner Channel 08896631190 21430 LFLC 8. fCAL turbo Partner Channel 08910367190 21490 FCAL |
| Classification | Class II |
| Recall number | Z-1476-2026 |
| Report date | Not disclosed in the FDA record. |
| Status | Ongoing |
| Reason for recall | Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel. |
What consumers should do
Facilities, clinicians, and consumers holding an affected unit or lot of cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001 cobas pro SSU, material number 09205632001 Affected Assays: (Product Name/ Catalog Number/ Application Code/ Application Short Name) 1. Cystatin C Gen.2 08105596190 20510 CYSC2 2. Ferritin Gen.4 08057648190 20571 FER4X 3. Lipoprotein (a) Gen.2 08106126190 20861 LPA2X 4. Lipoprotein (a) molarity 08106126160 20864 LPA2-X 5. Vancomycin Gen.3 08058849190 21211 VANC3O 6. Kappa Free Light Chains Partner Channel 08896640190 21421 KFLCX 7. Lambda Free Light Chains Partner Channel 08896631190 21430 LFLC 8. fCAL turbo Partner Channel 08910367190 21490 FCAL should stop using it for the purpose described in the recall notice and hold remaining stock pending instructions from Roche Diagnostics Operations, Inc.. FDA recall records typically direct the recalling firm to notify known accounts directly with a recall letter describing the specific corrective action -- replacement, field correction, refund, or return. Anyone who received the product through a distributor rather than directly from Roche Diagnostics Operations, Inc. should confirm with their supplier whether their specific lot or unit is on the affected list, since recall notices are lot- or unit-specific and not every unit of a product line is necessarily included. Questions about a specific unit's status should go to Roche Diagnostics Operations, Inc. directly or to the distributor that supplied it; Argus HQ is a public-record aggregator, not the recalling firm, and cannot confirm individual unit status.
What this classification means
A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote. FDA still requires the recalling firm to notify affected accounts and submit status reports on the correction, though the monitoring intensity is lower than for a Class I recall. FDA's Center for Devices and Radiological Health (for devices) or Center for Drug Evaluation and Research (for drugs) reviews the classification and can require additional corrective action if the initial response is judged insufficient. Because this record is filed as Class II, professionals and consumers handling an affected lot should treat FDA's own classification -- not this page's summary of it -- as the authoritative risk signal. A Class I recall is FDA’s most serious classification. FDA assigns Class I when there is a reasonable probability that use of, or exposure to, the recalled product will cause serious adverse health consequences or death. This classification does not mean harm has already occurred in every case -- it reflects FDA’s assessment of the realistic worst-case risk profile of the defect described in the recall, based on the reason for recall reported by the firm. Class I recalls receive the closest FDA monitoring of the three classes: the recalling firm typically must submit a recall strategy, periodic status reports, and effectiveness checks confirming that the affected lots have actually been removed from distribution or corrected in the field.
Roche Diagnostics Operations, Inc.’s FDA history
Argus HQ has recorded 6 total FDA actions tied to Roche Diagnostics Operations, Inc.: 0 warning letters, 6 recalls, 0 approval records, and 0 Form 483 inspection citations.
Frequently asked questions
- Where can I verify this recall directly with FDA?
- The verbatim FDA record is the source of truth for every recall Argus HQ tracks — always confirm against fda.gov before acting.
- Is cobas pro integrated solutions with cobas c 503 analytical u recalled?
- Yes. FDA lists a Class II recall by Roche Diagnostics Operations, Inc. for cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001 cobas pro SSU, material number 09205632001 Affected Assays: (Product Name/ Catalog Number/ Application Code/ Application Short Name) 1. Cystatin C Gen.2 08105596190 20510 CYSC2 2. Ferritin Gen.4 08057648190 20571 FER4X 3. Lipoprotein (a) Gen.2 08106126190 20861 LPA2X 4. Lipoprotein (a) molarity 08106126160 20864 LPA2-X 5. Vancomycin Gen.3 08058849190 21211 VANC3O 6. Kappa Free Light Chains Partner Channel 08896640190 21421 KFLCX 7. Lambda Free Light Chains Partner Channel 08896631190 21430 LFLC 8. fCAL turbo Partner Channel 08910367190 21490 FCAL, reported 20260311, with status "Ongoing." FDA recall number: Z-1476-2026.
- What is the recall class for Roche Diagnostics Operations, Inc.'s recall?
- FDA classified this recall as Class II. A Class II recall is FDA’s middle classification. FDA assigns Class II when use of, or exposure to, the recalled product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
- Why was cobas pro integrated solutions with cobas c 503 analytical u recalled?
- FDA's recall record states the reason as: Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.
Know if this happens again — we'll email you.
Free. No spam, just recall alerts for what you tell us to watch.
Related enforcement actions
Full FDA history for Roche Diagnostics Operations, Inc.- Recall: Roche Diagnostics Operations, Inc. — Class II recall: The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with s… →
- Recall: Roche Diagnostics Operations, Inc. — Class II recall: cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for u… →
- Recall: Roche Diagnostics Operations, Inc. — Class II recall: Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In v… →
- Recall: Roche Diagnostics Operations, Inc. — Class II recall: Tina-quant D-Dimer Test System, Fibrinogen/Fibrin Degradation Products Assay, Ma… →
- Recall: Roche Diagnostics Operations, Inc. — Class II recall: cobas infinity central lab, Material Number 07154003001 →
Compiled by Argus HQ Research from FDA primary sources · Reviewed by Andy Gaber, Founder
Cite this record
Reusing this data in reporting or research? Here’s a ready-made citation.
Argus HQ Research (2026). FDA Class II Recall: cobas pro integrated solutions with cobas c 503… by Roche Diagnostics Operations, Inc. — Recall Details. Digital Empire LLC. Retrieved from https://argushq.ai/recall/roche-diagnostics-operations-z-1476-2026-2026-03-11
"FDA Class II Recall: cobas pro integrated solutions with cobas c 503… by Roche Diagnostics Operations, Inc. — Recall Details." Argus HQ Research, Digital Empire LLC, 2026, argushq.ai/recall/roche-diagnostics-operations-z-1476-2026-2026-03-11.
Argus HQ Research. "FDA Class II Recall: cobas pro integrated solutions with cobas c 503… by Roche Diagnostics Operations, Inc. — Recall Details." Digital Empire LLC. Accessed July 17, 2026. https://argushq.ai/recall/roche-diagnostics-operations-z-1476-2026-2026-03-11.
@misc{argushq_argushq_ai_recall_roche_diagnostics_operations_z_1476_2026_2026_03_11_2026,
title = {FDA Class II Recall: cobas pro integrated solutions with cobas c 503… by Roche Diagnostics Operations, Inc. — Recall Details},
author = {{Argus HQ Research}},
year = {2026},
publisher = {Digital Empire LLC},
url = {https://argushq.ai/recall/roche-diagnostics-operations-z-1476-2026-2026-03-11},
note = {Accessed: July 17, 2026}
}Source: FDA.gov — Recalls ↗